Alteration of RPM speed rate in different ranges
Range
In steps of:
20 000 - 38 000
2 000 (e.g. 22 000, 24 000 etc....)
7. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
8. Press Program.
9. Press Torque to adjust the torque limit rate.
A pop-up window opens.
10. Adjust the torque limit rate.
The adjustment range is 10% - 95%.
11. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
12. Press Program.
13. Press Auto forward to adjust the counter-clockwise rotation time of the
micromotor in the Auto forward mode.
A pop-up window opens.
14. Adjust the Auto forward time.
The maximum time is 6 seconds and the minimum 0 seconds, and the
time can be changed in steps of 0.2 seconds.
15. Press OK.
The pop-up window closes and the new value is displayed on the control
panel.
25 Programming
182 Planmeca Compact i
User's manual