If necessary, the handpiece can be completely disassembled for more
thorough cleaning. The handpiece parts can be disinfected in a washer-
disinfector at 93°C, then optionally autoclaved at 134°C.
Saliva suction handpiece
The saliva suction handpiece can be disassembled for cleaning by pulling
out the end of the handpiece and removing the core (2) and the adjuster (1).
The handpiece parts can be disinfected in a washer-disinfector at 93°C, then
optionally autoclaved at 134°C.
28.6 Planmeca WEK water disinfection system
For information on how to use the Planmeca WEK water disinfection system,
see
Planmeca WEK water disinfection system user's manual.
28.7 Planmeca ProX
For information on how to clean the Planmeca ProX X-ray unit, see
Planmeca ProX user’s manual.
28 Cleaning and disinfection
238 Planmeca Compact i
User's manual