Direct simulation – current output *:
Using this function the requested value for the
current output can be defined.
For selecting the desired function within the
class please use the key
▲
or
▼
.
After selecting the function
Simulation dir-
ect current output
, press the ENTER key.
The actual value for the current will be dis-
played:
Set current
I = 12.50 mA
Example:
The current can be set within bounds of 0 to
23 mA. For changing the parameter please
refer to the description in Section "Input window
/ modify a value".
* This parameter is of relevance only, if the para-
meter
preset of
is set to „direct“.
8.1.10
Self-test
The function class SELF-TEST includes all
functions relating to the self-test of the sensor.
The diagnostic functions of the display electron-
ics, which monitor the signals of the sensor and
proper functioning of the electronics and the
software, are always active and cannot be
switched off.
Self-test on / off:
The
Self-test on/off
function is intended for
enabling or disabling the monitoring function of
the field coil current. This function is useful for
suppressing temperature dependencies of the
display electronics.
After selecting the function
Self-test on/off
using the key
▲
or
▼
press the ENTER key.
The current setting will be displayed.