Instruction Manual
DM08
One of the following settings can be chosen
("Selection window / make a selection", p. 25):
Setting
Function
direct
Status-, pulse-, und current out-
put* will be set directly
*
It is useful to make the presets for
the outputs before starting simula-
tion mode, using menu items
Simulation status output,
Simulation pulse output
and
Simulation current
output
.
In this manner the settings can be
modified specifically.
All outputs will be simulated simultan-
eously!
QVabs
The measuring variable will be set
Simulation preset Q*:
Using this function a „measuring variable” is
preset for simulating a flow rate. Flow rates for
both directions will be supported. The outputs
will be affected by the simulated value for the
measured value indirectly.
For selecting the desired function within the
class please use the keys
▲
and
▼
.
After selecting the function
Simulation pre-
set Q
, press the ENTER key to enter the para-
meter level for putting in the requested value:
Preset Q
+00900.00 m3/h
Example:
The simulation value is entered as described in
Section "Input window / modify a value".
* This parameter
Preset Q
is of relevance only, if
the parameter
Simulation preset Q /
direct
is set to "QVabs".
Direct simulation – status output *:
Using the function
Simulation direct
status output
the status output can be af-
fected directly.
For selecting the desired function within the
class please use the key
▲
or
▼
.
After selecting the function
Simulation dir-
ect status output
, press the ENTER key.
The current setting will be displayed:
Output
[off]
Example:
For changing the parameter please refer to the
description in Section "Selection window / make
a selection".
* This parameter is of relevance only, if the para-
meter
preset of
is set to „direct“.
Direct simulation – pulse output *:
Using the function
Simulation direct pulse
output
the requested frequency for the pulse
output can be defined.
For selecting the desired function within the
class please use the key
▲
or
▼
.
After selecting the function Simulation direct
pulse output, press the ENTER key. The cur-
rent value for the frequency will be displayed:
Set frequency
0210 Hz
Example:
The frequency can be set within bounds of 6 Hz
to 1100 Hz
For changing the parameter please refer to the
description in Section "Input window / modify a
value".
* This parameter is of relevance only, if the
parameter
preset of
is set to „direct“.