Instruction Manual
DM08
Each of them can be reset individually to the ini-
tial value 0.00.
In the first step the required totalizer has to be
chosen by using the
▲
or
▼
key.
After the confirmation with the ENTER key, the
function
Reset totalizer
must be selected
and the following will be displayed:
Reset total?
[yes]
To reset one of the totalizers, you need to
toggle to [yes] explicitly ("Selection window /
make a selection", p. 25).
By pressing Esc or toggling to [no] the current
action will be aborted without changing the to-
talizer.
8.1.4 Measurement processing
The MEASUREMENT PROCESSING func-
tional class includes all functions that affect the
processing of the measured values.
Before changing the current settings, the cus-
tomer password has to be entered.
Otherwise, the settings only can be watched but
not changed.
The ESC key is used for discarding changes.
Damping value:
The damping value
is intended to attenuate
abrupt changes of the flow rate or disturbances.
It affects as well the display of the measured
value as the current output and pulse output of
the DM08.
For selecting the desired function within a func-
tional class please use the key
▲
and
▼
.
By choosing the function
Damping value
and
pressing Enter, the parameter level is entered
and the current value for the damping is dis-
played:
Damping
03 s
Example:
The damping value can be varied within an in-
terval from 1 to 60s ("Input window / modify a
value", p. 25).
Notice
After a jump in the measuring variable
the output measured value reaches
about 99% of the new set point after 5
.
The factory setting is
3
seconds.
- 32 -