Installation, use and maintenance
8.2.1 DISPLAY OF CURRENT CALIBRATION FACTOR AND
RESTORING FACTORY FACTOR
CAL
CAL
CAL
By pressing the CAL key while the appliance is in Standby, the
display page appears showing the current calibration factor used.
If no calibration has ever been performed, or the factory setting
has been restored after previous calibrations, the following display
page will appear:
The word “Fact” abbreviation for “factory” shows that the factory
calibration factor is being used
If, on the other hand, calibrations have been made by the user, the
display page will appear showing the currently used calibration
factor ( in our example 0,998) .
The word “user” indicates a calibration factor set by the user is be-
ing used..
CAL
The flow chart alongside
shows the switchover logic
from one display page to
another
In this condition, the Reset
key permits switching from
User factor to Factory
factor.
To confirm the choice of
calibration factor, quickly
press CAL while “User” or
“Fact” are displayed.
After the restart cycle, the
K400 uses the calibration
factor that has just been
confirmed
CAL
CAL
CAL
CAL
RESET
RESET
TIME OUT
ATTENTION
When the Factory Factor is confirmed, the old User factor is deleted from the mem-
ory