Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 8

The PIECAL 434 is supplying the loop 
voltage. There is an control loop error. This 
may be a transmitter (set for upscale burnout) 
with a bad or missing sensor, or a short in the 
loop.

The PIECAL 434 is supplying the loop voltage. 
A calibrated transmitter is limiting the loop 
current to 12.00 mA. An additional 0.28 mA is 
not controlled by the transmitter and is leaking 
somewhere in the loop.

mA OUT, % OUT (Percent of 4 to 20 mA)

Find current leaks in loops caused by ground faults, moisture or corrosion. The 434 simultaneously 
supplies power to a 2 Wire Transmitter (or loop with a transmitter) while displaying the 4 to 20 mA 
output and the amount of current leaking in the loop.

1) Disconnect one or both input wires from 

the device to be calibrated.

2) Select 

“mA”

 or 

“% 4 to 20mA”

 with slide 

switch 

q

.

3)  Select 

“SOURCE”

 using slide switch 

w

.

4)  Turn the knob 

clockwise several times 

until full scale output (>24.000 
mA/125.00%) The display will indicate 

“PWR MEASURE”

 and “

LEAKAGE:

5)  Connect the red source lead of the PIECAL 

434 to the plus (+) input of the device and 
the black source lead to the minus (-).

The PIECAL 434 supplies a nominal 24 volts 
DC at 24 mA to the 2 Wire Transmitter or loop. The current passed by the transmitter will be 
accurately displayed by the PIECAL 434 along with an indication of leakage current at the bottom of 
the display. If there is an uncontrolled loop, a transmitter with upscale burnout and bad or missing 
sensor or a short the display shows “

i

LOOP > 24mA”

Note:

 Many installed transmitters will normally indicate 0.01 to 0.10 mA leakage without significant 

control problem. Unstable readings may indicate loose connections or the presence of moisture.

Using Ground Leak Detection

PWR MEASURE                 OUT

 

24.179

 mA

              

i

 LOOP > 24mA

PWR MEASURE                 OUT

 

12.280

 mA

    

LEAKAGE:     00.28mA

Enabling Ground Leak Detection

Double click the 

r

 DIAL KNOB at any time the unit is on and the following display will appear for 30 

seconds:
Turn the 

r

 DIAL KNOB to move through the menu. Press the 

r

 DIAL KNOB to toggle between OFF 

and ON or to change the steps setting. These settings are remembered even with the power off. 

EXIT MENU

 - exits this menu immediately and saves any changes. Menu will automatically exit after 

30 seconds of inactivity.

GROUND LEAK DETECTION

 - when ON the PIECAL 434 has the ability to check for current leaks 

caused by ground faults, moisture or corrosion that bypass the current control element or transmitter.

Setting Up Ground Leak Detection

Typical Error Conditions

Typical

2-Wire

Transmitter

+ IN - REF +OUT-

Transmitter Input

Sensor

Process Signal

Simulated Input 

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

 4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

12.280

PWR MEASURE  OUT

LEAKAGE:  00.28mA

mA

12.280

PROCESS

INDICATOR

www.

.com

[email protected]

1.800.561.8187

Содержание PIECAL 434

Страница 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Страница 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Страница 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Страница 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Страница 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Страница 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Страница 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Страница 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Страница 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Страница 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Страница 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Страница 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Отзывы: