PIE PIECAL 434 Скачать руководство пользователя страница 6

Practical Instrument Electronics

82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • [email protected] • www.piecal.com

Page 6

Signals exceeding ±60.00 VDC are indicated 
by 

i  

on the display.

Reading Milliamp Outputs

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

 4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

 4.340

                          READ

mA

Milliamp Output Signal

Controller

Transmitter

P/I

DCS

READ V

Choose this function to measure from -99.99 to +99.99V DC.
1) Select 

“VDC”

 with slide switch 

q

.

2) Connect the red (+) and black (-) leads of the PIECAL 434 

across the voltage source to be measured.

Any DC voltage from -60.00 to +60.00 volts may be 
measured. Loop power supplies, signal voltages at receivers, 
batteries and transmitter voltage drops may be measured. 
Signals exceeding ±60.00 VDC are indicated by 

i  

on the 

display. Signals exceeding ±99.99 VDC will be indicated by 
“OVERRANGE” flashing on the display.

Read DC Volts

Voltage to be Measured

Battery

Loop Power Supply

Transmitter Voltage Drop

1 to 5 Volt Loops

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

 4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

24.34

                          READ

V

              

 

 

READ 

i

         

               

72.83

 V

              

READ mA, READ % (Percent of 4 to 20 mA)

Choose this function to measure from 0.000 to 24.000 milliamps or -25.00 to 125.00%.

1) Open the current loop at any conve-

nient point along the signal path.

2)  Select “mA” or “% 4 to 20mA” with slide 

switch 

q

.

3)  Select “Read” using slide switch 

w

.

4) Connect the red input lead (+) of the 

PIECAL 434 to the more positive point 
of the break and the black input to the 
more negative point.

Signals below 0 mA or open circuits are 
indicated by 0.000 mA (-25.00%) on the 
display. Signals above 52 mA are current 
limited by protection circuitry and 
“OVERRANGE” is flashed on the display.

www.

.com

[email protected]

1.800.561.8187

Содержание PIECAL 434

Страница 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Страница 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Страница 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Страница 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Страница 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Страница 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Страница 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Страница 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Страница 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Страница 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Страница 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Страница 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Отзывы: