Source/Power & Measure Two Wire Transmitters

Ranges and Resolution

0.000 to 24.000 mA or -25.00 to 125.00% of 4-20 mA

Accuracy

≤ ± (0.01 % of R0.002 mA)

Noise

≤ ± ½ Least Significant Digit

Temperature effect

≤ ± 0.005 %/°C of FS

Loop compliance voltage

≥ 24 DCV @ 20.00mA

Loop drive capability - Leak Detection Off

Loop drive capability - Leak Detection On

1200 Ω at 20 mA for 15 hours nominal; 
950 Ω with Hart Resistor enabled
1000 Ω at 20 mA for 15 hours nominal; 
750 Ω with Hart Resistor enabled

Battery life

≥ 30 hrs at 12 mA nominal; ≥ 25 hrs with backlight on

2-Wire Transmitter Simulation

Accuracy

Same as Source/Power & Measure

Voltage burden

≤ 2V at 20 mA

Overload/Current limit protection

24 mA nominal

Loop voltage limits

2 to 60 VDC (fuse-less protected from reverse polarity 

connections)

Battery life

≥ 125 hours nominal; ≥ 100 hrs with backlight on

Voltage Read

Range and Resolution

-60.00 to +60.00 VDC Full Span (FS)

Accuracy

≤ ± 0.05 % of FS

Temperature effect

≤ ± 100 ppm/°C of FS

Input resistance

≥ 2 MΩ

Battery life

≥ 125 hours nominal; ≥ 100 hrs with backlight on

PIECAL 434 Specifications (continued)

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Accessories

INCLUDED:
 

Carrying Case with Belt Loop   

 

 

Part No. 020-0205

OPTIONAL:
 

AC ADAPTOR (200 to 240 VAC)  

 

 

Part No. 020-0100

 

AC ADAPTOR (100 to 120 VAC)   

 

 

Part No. 020-0101

 

Ni-MH 1 Hour Charger w/4 Ni-MH AA Batteries 

 

Part No. 020-0103

 

 

(100-120 V AC input for North America Only)

www.

.com

[email protected]

1.800.561.8187

Содержание PIECAL 434

Страница 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Страница 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Страница 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Страница 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Страница 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Страница 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Страница 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Страница 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Страница 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Страница 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Страница 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Страница 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Отзывы: