mA OUT, % OUT (Percent of 4 to 20 mA)

Choose this function to simultaneously supply power to a 2 Wire Transmitter while displaying the 4 
to 20 mA output of the transmitter.

1) Disconnect one or both input wires from 

the device to be calibrated.

2) Select 

“mA”

 or 

“% 4 to 20mA”

 with slide 

switch 

q

.

3) Select 

“SOURCE”

 using slide switch 

w

.

4) Press and turn the knob 

clockwise 

several times until full scale output (24.000 
mA/125.00%) is obtained. The display will 
indicate 

“PWR MEASURE”

.

5)  Connect the red source lead of the PIECAL 

434 to the plus (+) input of the device and 
the black source lead to the minus (-).

The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter. The current 
passed by the transmitter will be accurately displayed by the PIECAL 434. Calibrate the transmitter 
in the usual manner and disconnect the PIECAL 434. If Ground Leak Detection is enabled the 
amount of milliamps leaking in the transmitter or loop is indicated on the display.

Power & Measure 2-Wire Transmitters

Typical

2-Wire

Transmitter

+ IN - REF +OUT-

Transmitter Input

Sensor

Process Signal

Simulated Input 

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

 4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

OUT

%

25.42

www.

.com

[email protected]

1.800.561.8187

Содержание PIECAL 434

Страница 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Страница 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Страница 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Страница 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Страница 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Страница 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Страница 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Страница 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Страница 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Страница 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Страница 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Страница 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Отзывы: