2 Wire mA, 2 Wire % (Percent of 4 to 20 mA)

Choose this function to simulate a 2 Wire Transmitter output from 0.000 to 24.000 milliamps. 
Operates in loops with power supply voltages from 2 to 100 VDC

Simulate 2-Wire Transmitters

1) Disconnect one or both input wires 

from the device to be calibrated.

2) Select 

“mA”

 or 

“% 4 to 20mA” 

with 

slide switch 

q

.

3) Select 

“2 WIRE”

 using slide switch 

w

.

4) Connect the red input lead of the 

434 to the plus (+) input of the field 
connections and the black lead to 
the minus (-).

Loop current is adjusted in 0.001 mA (0.01%) increments by turning the knob 

r

 

while the 

EZ-CHECK™ switch 

e

 is in the “DIAL”  position, or the current can be set at the fixed points of 

4.000mA (0.00%) or 20.000mA (100.00%) with switch 

e

. Press and turn the knob for faster dialing 

with 0.100 (1.00%) increments. 

Start automatic stepping  or ramping by pressing and holding the 

r

 DIAL KNOB for 3 seconds. The 

word STEPPING or RAMPING will flash on the display anytime the selected automatic function is 
running. The PIECAL 434 will automatically step or ramp between 4mA & 20mA (or 20mA & 4mA) 
for 30 seconds then reverse direction. Stop the stepping or ramping by pressing or turning the knob 
or moving any switch.

To

Sensor

Typical

2-Wire

Transmitter

(Disconnected)

+ IN - REF +OUT-

Receiver

Powers External

2-Wire Transmitter 

Power Supply

(2 to 100 VDC) 

Power 2 Wire

Transmitters

Dial Output to

FULL SCALE to

Power 2 Wire

Transmitters

DIAL

AUTOMATED DIAGNOSTIC CALIBRATOR

 4-20 mA

Push & Hold

to

Store/Step

Push & Turn

for

Fast Dialing

Double Click

for Options

VDC

%mA

mA OFF

100%

0%

12.004

                        2 WIRE

mA

To change the Automatic Stepping & Ramping settings

Double click the 

r

 DIAL KNOB at any time the unit is on and the following display will appear for 30 

seconds:
Turn the 

r

 DIAL KNOB to move through the menu. Press the 

r

 DIAL KNOB to toggle between OFF 

and ON or to change the steps setting. These settings are remembered even with the power off. 

EXIT MENU

 - exits this menu immediately and saves any changes. Menu will automatically exit after 

30 seconds of inactivity.

STEPS

 - pressing the knob will cycle from 2, 3, 5 steps and RAMP. 

2 steps

 will automatically switch between 4 & 20 mA (0 & 100%).

 

3 steps

 between 4, 12 & 20 mA (0, 50 & 100%). 

5 steps

 between 4, 8, 12, 16 & 20 mA (0, 25, 50, 75 & 100%).

RAMP

 will continuously ramp the output between 4 & 20 mA (0 & 100%).

Setting Up Automatic Stepping & Ramping

www.

.com

[email protected]

1.800.561.8187

Содержание PIECAL 434

Страница 1: ...om 4 to 20 in 2 3 or 5 steps or choose a continuous ramp CALIBRATE USING LOOP POWER Check loop wiring and receivers by using the PIECAL 434 in place of a 2 Wire transmitter The PIECAL 434 uses any loo...

Страница 2: ...put 0 001 mA 0 01 at a time Push and turn the knob for faster dialing adjusting 0 100 mA 1 00 at a time Press and Hold the knob to start automatic stepping or ramping Double click the knob for steppin...

Страница 3: ...0 5 steps between 4 8 12 16 20 mA 0 25 50 75 100 RAMP will continuously ramp the output between 4 20 mA 0 100 GROUND LEAK DETECTION when ON the PIECAL 434 has the ability to check for current leaks ca...

Страница 4: ...4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch Sourcing Milliamps Power 2 Wire Transmitters Dial Output to...

Страница 5: ...l automatically step or ramp between 4mA 20mA or 20mA 4mA for 30 seconds then reverse direction Stop the stepping or ramping by pressing or turning the knob or moving any switch To Sensor Typical 2 Wi...

Страница 6: ...DC are indicated by ion the display Signals exceeding 99 99 VDC will be indicated by OVERRANGE flashing on the display Read DC Volts Voltage to be Measured Battery Loop Power Supply Transmitter Voltag...

Страница 7: ...e PIECAL 434 to the plus input of the device and the black source lead to the minus The PIECAL 434 supplies a nominal 24 volts DC at 24 mA to the 2 Wire Transmitter The current passed by the transmitt...

Страница 8: ...an indication of leakage current at the bottom of the display If there is an uncontrolled loop a transmitter with upscale burnout and bad or missing sensor or a short the display shows iLOOP 24mA Note...

Страница 9: ...E SIGNALS Signals below 0 mA or open circuits are indicated by 0 000 mA 25 00 on the display Signals above 24 mA are current limited by protection circuitry to approximately 25 mA OPEN LOOPS The displ...

Страница 10: ...t the 4 20 mA control wires from the Current to Pressure I P converter or the actuator 2 Connect the PIECAL 434 following the connection diagrams on the previous pages for Simulate 2 Wire Transmitters...

Страница 11: ...VAC for North America Only charger four NiMh batteries AC DC cords Part 020 0103 Low Battery Low battery indication with nominal 1 hour of operation left Protection against misconnection Over voltage...

Страница 12: ...Accuracy Same as Source Power Measure Voltage burden 2V at 20 mA Overload Current limit protection 24 mA nominal Loop voltage limits 2 to 60 VDC fuse less protected from reverse polarity connections...

Отзывы: