PicoSource PG900 Series User's Guide
43
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
6
Specifications and characteristics
The distinction between specifications, characteristics, typical performance, and
nominal values is as follows.
Specifications
describe guaranteed performance over the temperature range +15ºC
to +25ºC (unless otherwise noted). All specifications apply after the instrument’s
temperature has been stabilized after 1 hour of continuous operation. Unless otherwise
noted, corrected limits are given when specifications are achieved using software
processing.
Many performance parameters are enhanced through frequent, simple calibrations.
All specifications are subject to change without notice. Specifications are valid after a 1
hour warm-up period, and ±5°C from the firmware calibration temperature.
Characteristics
provide useful, but not guaranteed, information about the functions
and performance of the instrument.
Typical Performance
, where listed, is not guaranteed, but indicates the performance
that most units will exhibit.
Nominal Value
indicates the expected, but not guaranteed, value of the parameter.
Factory Calibration Cycle.
For optimum performance, the instrument should have a
complete verification of specifications once every 12 months.
6.1
PG900 family specifications
Specifications
PicoSource PG911
PicoSource PG912
PicoSource PG914
Generation
technique
Integrated step
recovery diode pulse
outputs
Drive outputs with
tunnel diode pulse
heads
Integrated step
recovery diode pulse
outputs and
drive outputs with
tunnel diode pulse
heads
Output
channels
2 SRD
2 TD
2 SRD + 2 TD
Output
waveform
Dual-channel positive
and negative pulse
generators with
controlled amplitude
and timing
parameters
Dual-channel positive
and negative pulse
generators with fixed
amplitude and
controlled timing
parameters
Four-channel positive
and negative pulse
generators with
controlled amplitude
and timing
parameters
Содержание PicoSource PG900 Series
Страница 2: ......