PicoSource PG900 software
34
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
External Trigger Menu
When
Ext Trigger
is selected from the
Period Source
menu the
External Trigger
menu opens.
External Trigger Menu
The
External Trigger
menu controls three settings:
Slope
,
Level
and
Holdoff
.
Note:
The input impedance on the
TRIGGER INPUT
connector is fixed at 50 Ω.
Slope
With the
Ext Trigger
period source selected, the positive edge or negative edge of the
source will trigger a pulse period.
Level
The
Level
sets the trigger level threshold in the range of –1 V to 1 V that will be
detected by the pulse generator as a valid trigger input pulse. Adjustable in coarse
increments of 10 mV or fine increments of 1 mV.
Holdoff
Holdoff is a count of a period of time. With external trigger it disables the trigger
circuit after a given period of time, after the last trigger occurred. Use holdoff to obtain
a stable trigger for HF repetitive, composite or complex waveforms.
Example 1.
External trigger frequency is much higher than generating pulse
frequency.
Example 2.
If the number or duration of sub-signals is known you can disable them
by choosing an appropriate holdoff value.
The
Holdoff
range is from 1 µs to 1 s in a 1-2-5-10 sequence. When fine mode is on,
you can change holdoff time in increments of up to 200 ns.