How to use your pulse generator
16
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
3.2
Free-running operation for SRD integrated pulse outputs
Integrated pulse outputs based on SRD technology are available with the PicoSource
PG911 and PG914 pulse generators. Follow the setup instructions below using the
PicoSource PG900 software.
1.
To switch on the positive SRD generator, click
Positive Out
in the
Output
menu, and then click
On
.
2.
Make sure that
Period Source
is set to
Int Clock
.
3.
Make sure the
Trigger
is set to
Continuous
.
4.
Set the
Period
in the
Timing
menu to provide the operation rate. Set the
desired period value from 1 µs to 1 s (1 MHz to 1 Hz).
5.
Set the
Width
in the
Timing
menu. Set the desired value from 200 ns to 4 s.
6.
Set the desired
Limit
of output amplitude, from 2.5 V to 6 V, to protect the
device under test. The change needs to be confirmed whenever the limit is
increased.
7.
Set the desired output
Amplitude
from 2.5 V to 6 V.