PicoSource PG900 Series User's Guide
47
Copyright © 2016 Pico Technology. All rights reserved.
pg900ug.en r1
6.5
External trigger input
Specifications
PicoSource PG911, PG912 and PG914
Bandwidth
DC to 1 GHz
Sensitivity
20 mV p-p DC to 100 MHz.
Increasing linearly from 20 mV p-p at 100 MHz to 100 mV p-p at
1 GHz.
Minimum
pulse width
500 ps @ 100 mV p-p
RMS trigger
jitter,
external
trigger input -
to - trigger
output
2.5 ps typ, 3 ps max
Trigger slope
Selectable rising or falling edge
Delay,
external
trigger input-
to-trigger
output
(4 ± 1) ns
Trigger level
range
–1 V to 1 V
Trigger level
resolution
1 mV
Rise time
TBD
Maximum safe
trigger input
voltage
16 dBm, or ±2 V (DC + AC pk)
Trigger
nominal input
impedance
(50 ± 0.5)
Ω
Trigger
coupling
DC-coupled
Trigger input
connector
SMA(f)
Trigger
holdoff
Range
1 µs to 1 s
Resolution
200 ns: 1 µs to 1.3 ms,
500 ns: > 1.3 ms to 5 ms,
1 µs: > 5 ms to 10 ms,
2 µs: > 10 ms to 20 ms,
5 µs: > 20 ms to 40 ms,
10 µs: > 40 ms to 85 ms,
20 µs: > 85 ms to 170 ms,
50 µs: > 170 ms to 1 s