Page 2.1
MATRIX MODULE 40-510/511/512/513
pickering
SECTION 2 - TECHNICAL DESCRIPTION
SECTION 2 - TECHNICAL DESCRIPTION
FUNCTIONAL DESCRIPTION
A functional block diagram is provided in Figure 2.1. The Matrix Module is powered by a +5V input via Compact PCI
connector J1. The interface to the user test equipment is via the front panel mounted 68-pin micro-D connector, J2.
A bank of relays, which are either single pole, screened single pole or two pole depending on the module type are
energised via control signals from storage registers U9 to U20. The storage registers are addressed by PCI bridge
U1 to output the required signal. Module configuration, i.e. 12 x 4, 12 x 8, 24 x 4, etc, is determined by hardwired
links and data stored in EEPROM U8. PCI Bridge U1 is configured by EEPROM U2.
Figure 2.1 - Matrix Module (40-510/511/512/513): Functional Block Diagram
68-PIN MICRO-D
CONNECTOR
J2
STORAGE
REGISTERS
U9 TO U20
MODULE
CONFIGURATION
U6, U7 AND U8
PCI BRIDGE
U1
PCI BRIDGE
CONFIGURATION
U2
TERMINATING
RESISTORS
R5 TO R16
COMPACT PCI
BUS
CONNECTOR
J1
+5V
0V
RELAY COILS
RELAY CONTACTS