5 Installation
N565 NEXACT® Linear Stage
MP148E
Version: 1.4.0
17
NOTICE
Excessively long screws!
Screws inserted too deeply can damage the N-565.
Pay attention to the depth of the mounting holes (p. 36).
Only use screws of the correct length for the respective mounting holes.
NOTICE
Protruding screw heads!
Protruding screw heads can damage the N-565.
Ensure that the screw heads do not protrude from countersunk holes so that they do not
interfere with the motion.
Requirements
You have read and understood the general notes on installation (p. 13).
You have provided a suitable underlying surface. For the required position of the holes,
see "Dimensions" (p. 36).
−
For mounting from
above with M3x12 screws
: Four M3 holes with a depth of at
least 5 mm are provided.
−
For mounting from
below with M4 screws
: Four through-holes are provided for M4
screws. The thickness of the underlying surface and the depth of the counterbores
for the through-holes in the underlying surface are matched so that compliance
with the maximum screw-in depth of 7.5 mm in the N-565 is adhered to.
−
The surface flatness is ≤10
µm.
−
For applications with large temperature fluctuations:
The surface should have the same or similar thermal expansion properties as the
N-565.
You have accounted for the space required to route cables according to regulations and
without bending them.
Tools and accessories
Screws according to the selected mounting option (see above):
−
Four M3x12 screws, included in the scope of delivery
or
−
Four M4 screws of suitable length, not included in the scope of delivery