Title
Service Manual, LPCR2
Page
46 of
104
Doc
Type
Service Product Drawing
Doc
ID
3323299
Doc
Rev
B
Doc
State
Released
State Date
11/14/2017
7000545 Revision A
Monthly Auto-Check-In
Overview
: This session explains how often and when the
CR2
will send device data to LIFELINK or LIFENET System
to update device status.
The
CR2
will send data when any of the following events occur:
1. Monthly Auto Check-In (initiated by the
CR2)
- The
CR2
automatically runs the monthly self-test and checks in
to LIFELINK or LIFENET to update device status
2. Manual check-in (initiated by a user) – replace the battery with a new battery
3. AED in use (lid opened)
4. AED event complete and lid closed
5. Self-Test failed (daily, weekly, monthly)
NOTE
:
1. Device status on LIFELINK or LIFENET is not a real time status. To conserve device battery, the
communication between device to LIFELINK or LIFENET will only happen in above five cases.
2. The Monthly Auto-Check-In date is set in each
CR2
during manufacturing. It cannot be changed. To view the
monthly Check-In date on the
CR2
, log in to LIFELINK and go to Equipment view.
3. In LIFENET, Next Expected Check-In Date/Time can be viewed from
Devices
View >
View Device Details