Title
Service Manual, LPCR2
Page
18 of
104
Doc
Type
Service Product Drawing
Doc
ID
3323299
Doc
Rev
B
Doc
State
Released
State Date
11/14/2017
7000545 Revision A
Shock Ready Time
Note:
These times apply for a device that has just reached a low
battery state (Readiness indicator has stopped flashing).
Fully automatic device:
The time to charge to 360 joules after electrode pads are applied
is 35 seconds or less.
The time to charge to 360 joules after power is turned on
is 45 seconds or less, if electrode pads are on the patient when
device is turned on.
Semi-automatic device:
The time to charge to 360 joules after electrode pads are applied
is 35 seconds or less.
The time to charge to 360 joules after power is turned on
is 45 seconds or less, if electrode pads are on the patient when
device is turned on.
cprCOACH Feedback Technology
Metronome rate
104 beats per minute
Setup options
Two options are available for CPR coaching:
•
Hands-Only (no rescue breaths)
•
30:2 compression to breath ratio
Communications
USB, Wireless 802.11 b/g/n, or Cellular data transfer to
LIFELINKcentral AED Program Manager or LIFENET System.
USB cable type 2.0 A Male to Micro-B is required for USB
communications.
Environmental
Note:
All performance specifications defined assume that the device
has been stored (two hours minimum) at the operating temperature
prior to operation.
Operating
Temperature
0° to 50°C (32° to 122°F)
Long-term Storage
Temperature
15° to 35°C (59° to 95°F)
Short-term Storage
Temperature
-30° to 60°C (-22° to 140°F) for a maximum of 1 week
Altitude
-382 to 4572 meters (-1253 to 15,000 feet) above sea level
Relative Humidity
5 to 95% (non-condensing)
Liquid and Solid
Ingress
IP55 per IEC 60529
Physical (with handle)
Height
9.7 cm (3.8 in)
Width
22.6 cm (8.9 in)
Depth
27.4 cm (10.8 in)
Weight
2.0 kg (4.5 lb) including electrodes and battery
Battery
Type
Lithium Manganese Dioxide (Li/MnO
2
), 12.0 V, 4.7 Ah, 55.8 Wh
Standby Life
4 years if installed in device that is not used
Weight
0.3 kg (0.7 lb)