9 Pixel Data Processing
Figure 9.21: LFSR (linear feedback shift register) test image
9.8.3
Troubleshooting using the LFSR
To control the quality of your complete imaging system enable the LFSR mode, set the camera
window to 512 x 512 pixels (x=0 and y=0) and check the histogram. If your frame grabber
application does not provide a real-time histogram, store the image and use a graphic software
tool to display the histogram.
In the LFSR (linear feedback shift register) mode the camera generates a constant
pseudo-random test pattern containing all grey levels. If the data transmission is correctly
received, the histogram of the image will be flat (Fig. 9.22). On the other hand, a non-flat
histogram (Fig. 9.23) indicates problems, that may be caused either by a defective camera, by
problems in the acquisition software or in the transmission path.
Figure 9.22: LFSR test pattern received at the frame grabber and typical histogram for error-free data
transmission
74 of 121
MAN076 09/2017 V1.0
Содержание MV3-D640I-M01-144-CL
Страница 10: ...1 Preface 10 of 121 MAN076 09 2017 V1 0...
Страница 34: ...4 Image Acquisition 34 of 121 MAN076 09 2017 V1 0...
Страница 46: ...6 I O Control 46 of 121 MAN076 09 2017 V1 0...
Страница 76: ...9 Pixel Data Processing 76 of 121 MAN076 09 2017 V1 0...
Страница 78: ...10 Thermoelectric Cooler TEC 78 of 121 MAN076 09 2017 V1 0...
Страница 82: ...11 Precautions 82 of 121 MAN076 09 2017 V1 0...
Страница 94: ...12 Hardware Interface 94 of 121 MAN076 09 2017 V1 0...
Страница 100: ...14 Troubleshooting 100 of 121 MAN076 09 2017 V1 0...
Страница 106: ...17 Support and Repair 106 of 121 MAN076 09 2017 V1 0...
Страница 108: ...18 References 108 of 121 MAN076 09 2017 V1 0...
Страница 112: ...A Pinouts 112 of 121 MAN076 09 2017 V1 0...
Страница 120: ...C Use Cases 120 of 121 MAN076 09 2017 V1 0...
Страница 121: ...D Document Revision History Revision Date Changes 1 0 July 2017 First version MAN076 09 2017 V1 0 121 of 121...