Error Messages
Chapter 10 - Troubleshooting
61
Error Messages
The following error messages are directly related to telemetry and appear
on the fetal monitor. Refer to the Instructions for Use provided with your
monitor for error messages not related to telemetry monitoring.
Series 50 Family
Message
Display
Cause
Solution
Err 9
US Toco
Invalid telemetry mode.
Check the cable from the telemetry
receiver and, if necessary, replace it.
Err 14
US Toco
Incorrect transducer
connected to transmitter.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Err 16
US Toco
Transducers are
connected to the front
panel of the monitor.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.