Solving General Problems
60
Chapter 10 - Troubleshooting
Transmission INOP
light on the receiver
is still lit when the
transmitter is
switched on.
Receiver and transmitter do not
have the same channel or serial
number.
Batteries in the transmitter are
exhausted.
Check channel number and the serial
numbers are the same on the receiver and
the transmitter.
Change the batteries in the transmitter
(refer to Chapter 3 for details.)
Battery Low Light lit
on receiver.
Power in batteries is low.
Change batteries.
INOP transmission
lamp is lit after the
patient has moved a
short distance away
form the receiver.
Local Antenna:
Antenna not connected correctly.
Remote Antenna System:
Antenna cable not connected
correctly to receiver.
Check antenna is connecting correctly.
Test the antenna system by bringing the
transmitter close to the receiver. If the
transmission is good, then the antenna
system is not functioning properly.
Contact your Philips Medical Response
Center.
Problem
Possible Causes
Solutions
Содержание 50 T Series
Страница 6: ...vi Contents ...
Страница 22: ...Application Overview 16 Chapter 2 Overview ...
Страница 40: ...Solving Common Monitoring Problems 34 Chapter 4 Monitoring ...
Страница 42: ...Solving Common Problems 36 Chapter 5 Fetal Movement Profile ...
Страница 46: ...Setting Up the Marker 40 Chapter 6 Marking Events ...
Страница 56: ...On a Wall 50 Chapter 8 Mounting the Receiver ...
Страница 74: ...Declaration of Conformity 68 Chapter 11 Accessories and Technical Specifications ...