![Philips 50 T Series Скачать руководство пользователя страница 68](http://html.mh-extra.com/html/philips/50-t-series/50-t-series_instructions-for-use-manual_319077068.webp)
Error Messages
62
Chapter 10 - Troubleshooting
8040A
8041A
Message
Display
Cause
Solution
Err 16
US Toco
Transducers are
connected to the front
panel of the monitor.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Message
Display
Cause
Solution
Signal
indicator
lamps
flashing
Indicator
Panel
Invalid telemetry mode.
Incorrect transducer
connected into
transmitter. (Only
Ultrasound and Toco
transducers can be used).
Transducers are
connected to the front
panel of the monitor.
Check the cable from the telemetry
receiver and, if necessary, replace it.
Check that the transducer is
compatible with Series 50T Fetal
Telemetry System.
Disconnect the transducers from
the monitor or switch off the
telemetry receiver.
Содержание 50 T Series