Testing the Parameter Signals
Chapter 9 - Maintenance and Performance Assurance
53
If the response is different, contact your Philips Medical Response
Center.
Signal
Correct Monitor Response
US
125
is displayed and printed.
Signal Quality Indicator is green.
Fetal heartbeat is heard from the loudspeaker.
Toco
A triangle signal with an amplitude of
40
units is
displayed and printed.
Each cycle lasts for 12 seconds.
DECG
150
is displayed and printed
Signal Quality Indicator is green.
Fetal heartbeat is heard from the loudspeaker.