17
User Manual
Alarm /
Alert
Priority
Audible
Indicator
Visual Indicators
Device Action
User Action
Patient
Disconnect
High
• • • • •
Red flashing button;
“Patient Disconnect”
Operates
This alarm occurs when the patient
circuit is disconnected or has a large
leak. Press the Alarm Silence/Indicator
button to silence the alarm. Reconnect
the patient circuit or fix the leak. If the
alarm continues, contact your home
care provider for service.
Low Input
Voltage
Medium
• • •
Yellow flashing button;
“Low Voltage”
Device Operates;
Humidifier shuts
down
The alarm is caused when input power
at the device, either from an AC outlet
or battery, falls below the acceptable
limit for 10 seconds.
Press the Alarm Silence/Indicator button
to silence the alarm. If the device is
plugged into a wall outlet, unplug the
device and then plug it back in. If the
alarm continues to occur, contact your
home care provider for service.
If you are using a battery, replace the
battery or plug the device into an AC
outlet. If the alarm continues, contact
your home care provider for service.
SD Card
Corrupted
Low
• •
Solid yellow button;
“SD card corrupted
Reformat card?”
Operates
This alarm occurs when a problem
exists with the SD card. The data may
be corrupted. Press the Alarm Silence/
Indicator button to silence the alarm.
Choose “yes” to reformat the card. If
you choose “no” the card will not be
reformatted. Note: Any information on
the card will be lost when reformatted.
Contact your home care provider with
any questions.
SD Card
Full
Low
• •
Solid yellow button;
“SD card full”
Operates
This alarm occurs when the SD card is
full. Press the Alarm Silence/Indicator
button to silence the alarm. Remove the
SD card and replace it.
SD Card:
Remove and
Reinsert
Low
• •
Solid yellow button;
“Reinsert SD Card”
Operates
This alarm occurs when the device
cannot read the SD card. The card may
be inserted incorrectly. Remove the SD
card and reinsert. If the alert continues
to occur, replace the SD card or contact
your home care provider.
SD Card:
Prescription
Rejected
Low
• •
Solid yellow button;
“SD card inserted:
prescription rejected”
Operates
This message occurs when the
prescription is missing or incorrect.
Contact your home care provider for
the correct prescription.