16
User Manual
Alarm /
Alert
Priority
Audible
Indicator
Visual Indicators
Device Action
User Action
Low
Pressure
High
• • • • •
Red flashing button;
“Low Pressure”
Operates
This could be caused by an excessive
leak or blockage or a device
malfunction. Press the Alarm Silence/
Indicator button to silence the alarm.
Remove your mask. Check for the
following: dirty inlet filters, blocked
air intake, excessive leak in the patient
circuit. If the alarm continues, contact
your home care provider.
Low Speed
High
• • • • •
Red flashing button;
“Low Blower Speed”
Operates
This may be caused by a malfunctioning
device. Press the Alarm Silence/Indicator
button to silence the alarm. Remove
your mask Remove power from the
device. Restore power. If the alarm
continues, contact your home care
provider for service.
High
Pressure
High
• • • • •
Red flashing button;
“High Pressure”
Operates;
If the alarm
continues for
10 seconds, the
alarm escalates
to a Device
Inoperative alarm
This may be caused by a malfunctioning
device. Press the Alarm Silence/Indicator
button to silence the alarm. Remove
your mask Remove power from the
device. Restore power. If the alarm
continues, contact your home care
provider for service.
Low
Pressure
Support
High
• • • • •
Red flashing button;
“Low Pressure
Support”
Operates
This could be caused by an excessive
leak or blockage or a device
malfunction. Press the Alarm Silence/
Indicator button to silence the alarm.
Remove your mask. Check for the
following: dirty inlet filters, blocked
air intake, excessive leak in the patient
circuit. If the alarm continues, contact
your home care provider.
Apnea
High
• • • • •
Red flashing button;
“Apnea”
Operates
This alarm is generated when an apnea
event occurs during therapy. Press
the Alarm Silence/Indicator button to
silence the alarm. Report the alarm to
your home care provider. Continue
using your device.
Low Minute
Ventilation
High
• • • • •
Red flashing button;
“Low Minute Vent”
Operates
This alarm occurs when the calculated
minute ventilation is less than or equal
to the alarm setting. Press the Alarm
Silence/Indicator button to silence the
alarm. Report the alarm to your home
care provider. Continue using your
device.