11
User Manual
Setup Screen
From the Home screen, highlight “Setup” and press the wheel. The following Setup screen will appear. The user can
change settings in the Setup menu.
Se
tu
p
Back
Bi-Flex
1 2 3
Rise time
0 1 2 3
Ramp time
0:00 - 0:45
Ramp start
4 - EPAP Min
Tubing type
15 22 15H
SYSTEM ONE resistance
X1 X2 X3 X4 X5
Heated tube humidification on off
Humidity level
1 2 3
Tube temperature
0 1 2 3 4 5
SYSTEM ONE humidification on off
Humidifier
0 1 2 3 4 5
Auto on
on off
Humidifier LED Backlight
on off
Language
EN ES
Back
Setup Screen
note:
The screen will only show a few lines at a time. As you rotate the wheel to toggle over different options
the screen will slide up and down accordingly. If the text is too long to completely fit on the screen, it will scroll
horizontally across the screen when highlighted.
• Bi-Flex - The Bi-Flex comfort feature allows you to adjust the level of air pressure relief that you feel when you
exhale during therapy. Your home care provider can enable, lock or disable this feature. When your provider enables
Bi-Flex, a level will already be set for you on the device. If this is not comfortable, you can increase or decrease the
setting. The setting of “1” provides a small amount of pressure relief, with higher numbers providing additional relief. If
the provider has disabled this feature, this setting will not display.
note:
This same setting is also available under the “Flex/Rise time” screen.
• Rise time - The Rise time comfort feature is the time it takes for the device to change from EPAP to IPAP. If
Rise time is prescribed for you, you can adjust the Rise time from 0 to 3 to find the setting that provides you with
the most comfort. A setting of 0 is the fastest Rise time, while 3 is the slowest. This setting will not display if your
provider has not enabled Rise time on your device.
note:
This same setting is also available under the “Flex/Rise time” screen.
• Ramp time - This enables you to modify the Ramp time setting in 5 minute increments. The range for this setting
is 0 to 45 minutes.
• Ramp start - This displays the Ramp starting pressure. You can increase or decrease the Ramp starting pressure
from 4 cm H
2
O to the minimum EPAP setting in 0.5 cm H
2
O increments. This will only display if your provider has
enabled Ramp on your device.
• Tubing Type - This setting allows you to select the correct size diameter tubing that you are using with the device.
You can choose either (22) for the Philips Respironics 22 mm tubing, or (15) for the Philips Respironics 15 mm tubing.
When using the heated tube with the humidifier, the device will automatically change this setting to the appropriate
tubing type (15H) and you will not be able to change it. If your provider has locked this setting into place, you can
view the setting but cannot change it, and the screen will display a lock symbol ( ) next to the setting.
note:
If the Heated Tubing is removed, the device will default back to the previous tubing type setting.
• SYSTEM ONE resistance
(
)
- This setting allows you to adjust the level of air pressure relief based on
the specific Philips Respironics mask. Each Philips Respironics mask may have a “
System one
” resistance control
setting. Contact your home care provider if you cannot find this resistance setting for your mask. If your provider
has locked the resistance setting into place, you can view the setting but cannot change it, and the screen will
display a lock symbol ( ) next to the setting. If your provider has disabled resistance, you will not see this setting.