10
User Manual
Mask Fit Check Feature
The Mask Fit Check feature can be enabled or disabled by your home care provider. This feature allows you to check
the fit of your mask prior to starting therapy. This is done by measuring the amount of leak.
Put on your mask assembly. If Mask Fit Check is enabled, before you turn on the airflow, press the RAMP (
)
button on the top of the device. Airflow will start and the Mask Fit Check screen will appear, shown below.
Mask Fit Check
The device will deliver a test pressure while the screen will count down 40 seconds. After the test, normal therapy
will start and the screen will either display a checkmark (
) or an . The
shows that the leak found allows for
optimal performance of the device. The shows that the leak may affect device performance, however, the device
will remain functional and deliver therapy.
note:
If you choose to try to improve your mask fit, you can stop therapy, adjust the fit of your mask, and rerun the
Mask Fit Check. Please refer to the instructions that came with your mask and headgear for the proper fitting procedure.
Humidifier Preheat
When using a humidifier, the device can preheat the water tank for up to 30 minutes prior to starting therapy.
In order to activate the preheat mode, the blower must be “off” and a humidifier must be attached. From the device
Home screen, highlight “Therapy”, then press and hold down the control wheel for 5 seconds. You will hear a single
beep and the device will now be in preheat mode. The humidifier icon (
) will illuminate during this time.
During the 30 minute preheat, you will still be able to use the control wheel to select other menu options from the
Home screen. If you press the wheel while “Therapy” is highlighted on the Home screen, preheat mode will end and
the blower will turn “on” to begin therapy. The humidifier number selected in the setup menu (0, 1, 2, 3, 4, or 5) will
now take effect.
Bi-Flex/Rise time Screen
From the Home screen, highlight “Bi-Flex” or “Rise time” and press the wheel. The following screen will appear.
note:
This screen will be blank on the Home screen if your provider has not enabled Bi-Flex or Rise time on your device.
Back
Bi-Flex
1 2 3
Rise time 0 1 2 3
Back
Flex Screen
• Bi-Flex - The Bi-Flex comfort feature allows you to adjust the level of air pressure relief that you feel when you
exhale during therapy. Your home care provider can enable, lock or disable this feature. When your provider
enables Bi-Flex, a level will already be set for you on the device. If this is not comfortable, you can increase or
decrease the setting. The setting of “1” provides a small amount of pressure relief, with higher numbers providing
additional relief. If the provider has disabled this feature, this setting will not display.
note:
This same setting is also available under the “Setup” screen.
• Rise time - The Rise time comfort feature is the time it takes for the device to change from EPAP to IPAP. If
Rise time is prescribed for you, you can adjust the Rise time from 0 to 3 to find the setting that provides you with
the most comfort. A setting of 0 is the fastest Rise time, while 3 is the slowest. This setting will not display if your
provider has not enabled Rise time on your device.
note:
This same setting is also available under the “Setup” screen.