46
SPECIFICATIONS
Functional specifications
Product name
Pharmaceutical refrigerator
MPR-715F
Temperature controller
Electronic control system
Temperature display
Digital display
Thermal sensor
Thermistor sensor
Alarm temperature
Chamber set temperature ±2
o
C to ±14
o
C (refrigerator)
Chamber set temperature ±5
o
C to ±15
o
C (freezer)
Buzzer suspended period
10, 20, 30, 40, 50, 60 minutes, no recovery
Door alarm
2 minutes after door opening
Lock of chamber set temperature
ON or OFF
Alarms
High temperature alarm, Low temperature alarm, Door alarm
0
o
C alarm (refrigerator only)
Power failure alarm (with MPR-48B1 installed)
Self diagnostic
Sensor abnormality, Abnormal temp. of cooling circuit
Fan motor life, Battery life (with MPR-48B1 installed)
Remote alarm terminal
Allowable contact capacity: DC 30 V, 2 A
*
Back-up of setting
Nonvolatile memory
Specifications will be subject to change without notice.
*
It is recommended to use cables with a maximum length of 30 meters.
Performance specifications
Product name
Pharmaceutical refrigerator
MPR-715F
Model number
MPR-715F-PA
MPR-715F-PE
Maximum cooling performance
Refrigerator
Chamber center temperature: 2
o
C (ambient temp.; 35
o
C, no load)
Freezer
Chamber center temperature: -30
o
C (ambient temp.; 30
o
C, no load)
Temperature control range
Refrigerator Between
2
o
C and 14
o
C (ambient temp.; -5
o
C to 30
o
C, no load)
Freezer Between
-30
o
C and -20
o
C (ambient temp.; -5
o
C to 30
o
C, no load)
Rated voltage
AC 115 V
AC 220 V / 230 V / 240 V
Rated frequency
60 Hz
50 Hz
Power consumption
490 W *
275 W / 280 W / 285 W
Amperage
5.2 A *
2.1 A / 2.2 A /2.3 A
Maximum heat radiation
1764 kJ/h
1584 kJ/h /1620 kJ/h /1656 kJ/h
Noise level
43 dB
Maximum pressure
1900 kPa
Usable condition
-5
o
C to 30
o
C, 80 %R.H. or less
Specifications will be subject to change without notice.
The above data is measured based on our internal basis.
The unit with CE mark complies with EU directives.
* These values are maximum amperage and power consumption.