SETTING OF ALARM TEMPERATURE
ETTING OF
Setting of high temperature alarm (freezer)
The abnormality (chamber temperature rise) is noticed by blinking of alarm indicator and chamber
temperature display and sounding of the buzzer (15 minutes after blinking) if the chamber temperature
exceeds the set value of high temperature alarm. Always set the high temperature alarm to protect the
storage items from damage resulting from temperature rise.
Setting range of work temperature for high temperature alarm:
Between 5
o
C and 15
o
C higher than the chamber set temperature
Initial setting (factory setting): 10
o
C higher than the chamber set temperature
Example: Change the alarm temperature to chamber set temperature plus 5
o
C from chamber
set temperature plus 10
o
C
Details: Change from ”the setting that the high temperature alarm works at 10
o
C higher than the chamber set
temperature” to ”the setting that the high temperature alarm works at 5
o
C higher than the chamber set
temperature.”
•
Following is an example. Operate the key according to the desired alarm temperature.
1.
Depress the temperature display changeover key to light the indicator
for freezer temperature.
►
The current freezer temperature is displayed on the temperature display.
2.
Depress the up arrow key for 5 seconds.
►
The current chamber temperature on the temperature display changes to
function code F00 and the first figure (0) blinks.
3.
Depress the up arrow key 3 times.
►
The display changes to F03 from F00.
The figure changes between 0 and 9 by depressing the up arrow key.
4.
Depress the set key once.
►
The current setting (010) is displayed and the first figure (0) blinks.
The initial setting 010 (low temperature alarm works when the chamber
temperature is 10
o
C higher than the chamber set temperature) is displayed at the
first setting after installation.
5.
Depress the up arrow key 5 times (as needed for change).
►
The display changes to 015 from 010.
The figure changes between 0 and 9 by depressing the up arrow key.
6.
Depress the scroll key once.
►
The second figure starts to blink.
The first figure or second figure blinks alternately by depressing the scroll key.
7.
Depress the up arrow key once.
►
The display changes to 005 from 015.
The figure changes between 0 and 1 by depressing the up arrow key.
8.
Depress the set key once.
►
The work temperature of high temperature alarm is memorized and the current
chamber temperature is displayed on the temperature display.
The setting mode returns to the temperature display mode automatically when 90 seconds has passed without
any key operation (auto-return function). In this case, the setting is not accepted.
<Important> Never input any function code other than the specified one.
Change of first figure
Change of second figure
First figure is changeable
First figure is changeable
Change of first figure
Enter of the setting
Second figure is changeable