44
TEMPERATURE RECORDER
(OPTION)
The chamber temperature can be recorded and checked by installing an optional temperature recorder
(MTR-G3504, MTR-0621LH, or MTR-4015LH).
Contact our sales representative or agent for the purchase of temperature recorder.
Main specifications of temperature recorder
MTR-G3504 MTR-0621LH
MTR-4015LH
Application Refrigerator
&
Freezer Refrigerator
Freezer
Recording
range
-60
o
C to +40
o
C, -10
o
C to +40
o
C -6
o
C to +20
o
C -40
o
C to +14
o
C
Feed speed of
recorder chart
1-day/1 turn, 7-day/1 turn,
32-day/1 turn
31-day/batch 31-day/batch
Recorder chart
Circular type
Strip type
Strip type
Power source
Supplied from the refrigerator
Dry cell
Dry cell
For the installation of temperature recorder MTR-G3504, optional recorder mounting bracket MPR-S7 is
necessary.
For the installation of temperature recorder MTR-0621LH or MTR-4015LH, optional recorder mounting bracket
MPR-S30 is necessary.
BATTERY MOUNTING BOX
(OPTION)
It is recommended to install an optional battery mounting box (MPR-48B1) to prevent the rise of
chamber temperature in the case of power failure. At the power failure, the alarm activates (blink of
alarm indicator, sound of buzzer) to prompt an action for preventing the storage items.
Contact our sales representative or agent for the purchase of battery mounting box.