INSTALLATION LOCATION
This unit must be installed in a location which meets all the conditions described below.
If the unit is installed in a location which does not meet the conditions, its specified performance may not be
achieved or malfunctioning and accidents may occur.
A location not exposed to direct sunlight
Avoid any location which is exposed to direct sunlight. Installing the unit in a location exposed to
direct sunlight may cause its cooling performance to be reduced.
A well-ventilated (airy) location
In order to ensure ventilation, leave clearances of at least 10 cm around the unit (at the left, right and
back). Blocking the ventilation may reduce the unit’s cooling performance or cause malfunctioning.
A location away from sources of heat
Avoid any location which is close to a large source of heat (such as a heater or boiler). Installing the
unit near a large source of heat may reduce the unit’s cooling performance.
A location with minimal changes in temperature
Avoid any location where the ambient temperature is subject to sudden changes. If the unit is
installed in a location where the ambient temperature is subject to sudden changes, it will not be
possible to achieve a stable cooling performance.
A level location where the floor is capable of bearing the total combined weight (p optional
acces storage items)
Install the unit in a location which is not uneven and which is capable of bearing the total combined
weight (p optional acces storage items). If the unit is installed where the surface is
uneven or where the unit will be inclined at an angle, the unit will be unstable, and accidents or injuries
may result and/or vibration or noise may be generated.
A location free of ignitable or corrosive gases
Avoid any location exposed to ignitable or corrosive gases. Ignitable or corrosive gases can cause
explosions and/or a fire. Furthermore, corrosion of the electrical parts may cause the insulation to be
reduced and result in ground faults and/or electric shock.
CAUTION
Do not install the unit in any location (such as near a
waste-water facility) where sulfur compounds and other
corrosion-causing substances may be generated.
Corrosion of the copper pipes will cause the cooling unit to
deteriorate, possibly causing the product to malfunction.
A location with minimal humidity
Install the unit in a location where the relative humidity is less than 80 %R.H. Installing the unit in a
very humid location may cause ground faults and/or electric shock.