HANDLING OF STORAGE ITEMS
Always follow the indication below when taking out/putting in the material.
Put on the gloves when taking out/putting in the storage items in the freezer chamber. No gloves
may cause injury.
Do not put the excessive storage items in the chamber. Over putting disturbs the cold air
circulation and degrades the cooling performance.
Open/close the doors quickly or the chamber temperature rises.
Do not put a large amount of storage items of high temperature at a time.
Put the storage items close to the shelf center to avoid the contact with the chamber wall. Keep an
adequate space between materials. Contacting with chamber wall or other materials causes
uneven cooling in the chamber.
INITIAL SETTING, SETTING RANGE, FUNCTION CODE
Following shows the initial setting (factory setting) and setting range of each item. Change the setting
as needed.
Refer to the each related page for the details and procedure of the setting.
Setting item
Initial setting
Setting range (figure)
Function
code
Reference
page
Refrigerator temp.
5
o
C 2
o
C to 14
o
C (none) 21
Freezer temp.
-30
o
C -35
o
C to -15
o
C (none) 21
High temp. alarm
Refrigerator
5
o
C higher than the
chamber set temp.
between 2
o
C and 14
o
C
higher than the chamber
set temperature
F01 26
Freezer
10
o
C higher than the
chamber set temp.
between 5
o
C and 15
o
C
higher than the chamber
set temperature
F03 28
Low temp. alarm
Refrigerator
5
o
C lower than the
chamber set temp.
between 2
o
C and 14
o
C
lower than the chamber
set temperature
F02 27
Freezer
10
o
C lower than the
chamber set temp.
between 5
o
C and 15
o
C
lower than the chamber
set temperature
F04 29
Buzzer suspended
period
30 minutes
10, 20, 30, 40, 50, 60
min., or no recovery
F25 30
Lock of chamber
temperature setting
OFF
ON or OFF
(none)
22
The enter of function code through the up arrow key and scroll key is needed to set the alarm
temperature, recovery time, delay time. For the procedure, see the reference page.
<Important> Never enter the function code other than specified one at the time of setting. The
other function code is for service personnel only.