70Xi Cold Flow Properties Laboratory Analyzer
User Guide
24
1.
Viscosity can be calibrated by using jet fuel standards with known KV values. It is recommended
to test at least two jet fuel standards with different viscosity. Each standard should be tested
twice for repeatability.
2.
Press the ANALYZER OPTIONS
button from the Home screen. Then press PERFORM
NEW VISCOSITY CALIBRATION
2. Enter the Sample ID and expected KV for the first calibration standard. Press START CALIBRATION
TEST. The analyzer will prompt you to load the sample into a vial, place it in the sample inlet port, and
press OK.
2
H o w t o C a l i b r a t e V i s c o s i t y