70Xi Cold Flow Properties Laboratory Analyzer
User Guide
17
1.
There are 3 different user levels that allow varying degree of configuration control:
a.
User: lowest level that may have restricted access to certain features. This is the
default user level upon power up
b.
Administrator: level that can be password protected and has access to all available
features. Administrators can also configure access permissions of certain features for the
User group
c.
Factory: reserved for factory use only
2.
To change the current user level, press the ACCESS button
at the lower left corner. The
“Access” screen appears (see Fig. below)
3.
Select either USER or ADMINISTRATOR
4.
To setup or modify the password at the Administrator level:
a.
Switch user level to Administrator
b.
From the Home screen, press the ANALYZER OPTIONS button
.
H o w t o C h a n g e
U s e r L e v e l A c c e s s