3.4

 

DISPLAY

 

MODES

 

There are three measuring interface display modes: normal mode, A‐scan mode and B‐scan mode. 

And there are three display modes of Normal interface: The thickness value mode, Difference/rate of 
reduction  measurement  mode,  MAX/MIN  measurement  mode.  Select  this  in  “VIEW  MODE”  of 
configuration. 

ATTENTION:

 When the probe is properly “coupled” to the surface of the object being tested, the 

display will be shown in WHITE. If not coupled properly then the values will be shown in GREEN. When 
the values exceed the upper or lower limit setting then the values will be shown in RED. 

 

THICKNESS

 

VALUE

 

MODE:

 This interface mainly shows the velocity of material, the present thickness 

value and unit.     

 
 
 
 

 
 
 
 
 

                3.4 NORMAL MODE INTERFACE 

1—the

 

present

 

thickness

 

value

     

2—unit

     

3—material

 

velocity

     

4—battery

 

power

   

5—A—scan

 

interface

 

 

 

DIFFERENCE/RATE

OF

REDUCTION

 

MODE:

  This  interface  shows  the  measured  value,  the  preset 

nominal thickness value, the difference between the measured value and the preset value and the ratio 
between the difference and the normal value. Before using this mode, presetting the nominal thickness 
is needed. The method can be taken according to chapter 3.6. 
 
 
 
 
 
 
 
 
   

Содержание UTG-4000

Страница 1: ...l No UTG 4000 www phase2plus com 283 Veterans Blvd Carlstadt NJ 07072 201 933 6300 www phase2plus com 21 Industrial Ave Upper Saddle River NJ 07458 P 201 962 7373 F 201 962 8353 E mail info phase2plus...

Страница 2: ...D FUCTIONS 5 3 MEASURING THE THICKNESS 6 3 1 PRESET THE MATERIAL S VELOCITY 6 3 2 PRESET OTHER SPECIFICATIONS 7 3 3 CALIBRATION 8 3 4 DISPLAY MODES 9 3 5 ADJUSTING THE SPECIFICATIONS IN A SCAN MODE 11...

Страница 3: ...wn thickness or sound velocity 1 1 CONSTRUCTION OF THE GAGE 1 2 STANDARD CONFIGURATION Qty Thickness Gauge 1 5MHz PROBE 1 AA Batteries 2 2oz Couplant Gel 1 Rugged Carry case 1 Operation Manual 1 1 3 O...

Страница 4: ...Error 0 001 up to 0 984 0 007 up to 3 03 0 019 4 and above Display Modes Digital Thickness Readout A scan or B scan Min Max Capture Diff Value Reduction V Path Correction Automatic Update Rate Selecta...

Страница 5: ...mode getting the difference between the actual value and the normal value as well as the percentage 8 Inch Metric Conversion 9 Extra large memory up to 100 000 values 1000 waveforms 10 Selectable Reso...

Страница 6: ...may be different from the actual material velocity intended to be measured Calibration to a known thickness Note This procedure requires a sample piece of the specific material to be measured the exa...

Страница 7: ...ctual thickness is less than the minimum value preset MAXIMUM ALARM set the maximum thickness alarm value The result will be displayed in red if the actual thickness is more than the maximum value pre...

Страница 8: ...TION STEP ILLUSTRATION 1 Adjust the sound velocity to 5900m s according to the method in chapter 3 1 2 After turning on the gage press to enter the calibration interface The screen shows PUT PROBE ON...

Страница 9: ...in GREEN When the values exceed the upper or lower limit setting then the values will be shown in RED THICKNESS VALUE MODE This interface mainly shows the velocity of material the present thickness va...

Страница 10: ...ITS VALUE SCANNING MODE this mode allows the customer to test the thickness of material continuously and to show the upper lower limits after the tests It shows the minimum and maximum values during t...

Страница 11: ...robe and the object are not completely coupled the letters in the various interfaces are shown in GREEN when properly coupled they are displayed in WHITE color and when the either the upper or lower l...

Страница 12: ...this will only show when the GATE specification is highlighted The first intersection point between the waveform and gate is shown with a RED arrow which will help you determine if this thickness valu...

Страница 13: ...n be realized by adjusting the gate setting above the defect echoes 3 9 REAL CASE 2 3 If there are some surface faults in the testing object and the gate locks the defect echoes the measuring result w...

Страница 14: ...ty 10 The minimum thickness value on the B scan image 11 Parameter display area 12 Gain value 3 72 Introduction to B scan The UTG 4000 has a time base B scan function By moving the probe across the su...

Страница 15: ...icates the length of echo blanking the waveform above this is considered invalid See fig 3 16 below 1 Identification of dual echo mode 2 Blue arrow indicates secondary echo 3 E blanking 4 Blue line th...

Страница 16: ...ferred to your PC via USB output and saved as an EXCEL or TXT file 4 1 THICKNESS VALUE AND WAVEFORM STORAGE 1 Storage file number 2 Line mark 3 Row mark 4 Data in memory 5 Return to main menu 6 Save c...

Страница 17: ...l attention to polarity Attention when not using the gauge for extended periods of time please remove batteries to prevent any leakage or corrosion 5 2Maintenance 1 Always keep your gauge probe cable...

Страница 18: ...t or near the temperature of the material to be measured This will allow the gauge to correctly calculate the velocity of sound through the hot material When performing measurements on hot surfaces it...

Страница 19: ...d in ultrasonic thickness measurement does not travel through air efficiently A wide variety of couplant materials may be used in ultrasonic gauging Propylene glycol is suitable for most applications...

Страница 20: ...n 0 230 5900 Iron cast 0 180 4600 Lead 0 085 2200 Magnesium 0 230 5800 Mercury 0 057 1400 Molybdenum 0 250 6300 Monel 0 210 5400 Neoprene 0 063 1600 Nickel 0 220 5600 Nylon 6 6 0 100 2600 Oil SAE 30 0...

Страница 21: ...t Beijing 100192 China Tel 86 10 59792409 Fax 86 10 59814851 General E mail info phase2china com cn MEXICO Phase II de Mexico Calle A No 4 Promer Piso Col San Marcos Azcapotzalco C P 02020 Mexico Tel...

Отзывы: