Err117
Excess temperature pump bottom
part
– Cooling deficient
Optimize cooling
Check the ambient conditions
Err118
Excess temperature power stage
– Cooling deficient
Optimize cooling
Check the ambient conditions
Err119
Excess temperature bearing
– Cooling deficient
Optimize cooling
Check the ambient conditions
Err143
Excess temperature operating fluid
pump
– Cooling deficient
Optimize cooling
Check the ambient conditions
Reset at rotation speed f = 0 only
Err777
Nominal rotation speed not con-
firmed
– Nominal rotation speed not con-
firmed after replacement of the elec-
tronic drive unit
Confirm the nominal rotation speed via
[P:777]
Reset at rotation speed f = 0 only
Wrn007 Low voltage / mains power failure
– Mains failure
Check mains supply
Wrn018 Remote priority conflict
– Pumping station is switched on via
[P:010]
, whereas the E74-input "start/
stop" is off (opened)
Switch on the pumping station via E74
Switch off
[P:010]
Wrn045 High temperature motor
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn076 High temperature electronic
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn097 Pump information invalid
– Pump data faulty
Reset for default values
Wrn098 Pump information incomplete
– Connection to the pump is faulty
Contact Pfeiffer Vacuum Service
Wrn113
Rotor temperature inaccurate
– Measurement not calibrated
Contact Pfeiffer Vacuum Service
Wrn115
Temperature analysis pump bottm
part faulty
Contact Pfeiffer Vacuum Service
Wrn116
Temperature analysis bearing faulty
Contact Pfeiffer Vacuum Service
Wrn117
High temperature pump bottom part
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn118
High temperature power stage
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn119
High temperature bearing
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn143 High temperature operating fluid
pump
– Cooling deficient
Optimize cooling
Check the ambient conditions
Wrn168 High deceleration
– Rate of pressure rise too high; Venting
rate to high
Check and optimize the venting rate
(pump specific)
Error
code
Problem
Possible cause
Remedy