![Pfeiffer TC400 Скачать руководство пользователя страница 17](http://html.mh-extra.com/html/pfeiffer/tc400/tc400_operating-instructions-manual_1534853017.webp)
17
The Pfeiffer Vacuum parameter set
Set value settings
7.3
Configuring the connections
The electronic drive unit is pre-configured in the factory. Thereby the turbopump
is immediately operational with the necessary functions. The connections of the
electronic drive unit can be configured to suit individual requirements using the
parameter set.
Accessory connection
Configuration via parameters
[P:035]
,
[P:036]
,
[P:037]
or
[P:038]
.
326
TempElec
Temperature electronic
1
R
°C
0
999999
330
TempPmpBot
Temperature pump bottom part
1
R
°C
0
999999
336
AccelDecel
Acceleration / Deceleration
1
R
rpm/s
0
999999
342
TempBearng
Temperature bearing
1
R
°C
0
999999
346
TempMotor
Temperature motor
1
R
°C
0
999999
349
ElecName
Name of electronic drive unit
4
R
354
HW Version
Hardware version electronic drive unit
4
R
360
ErrHist1
Error code history, pos. 1
4
R
x
361
ErrHist2
Error code history, pos. 2
4
R
x
362
ErrHist3
Error code history, pos. 3
4
R
x
363
ErrHist4
Error code history, pos. 4
4
R
x
364
ErrHist5
Error code history, pos. 5
4
R
x
365
ErrHist6
Error code history, pos. 6
4
R
x
366
ErrHist7
Error code history, pos. 7
4
R
x
367
ErrHist8
Error code history, pos. 8
4
R
x
368
ErrHist9
Error code history, pos. 9
4
R
x
369
ErrHist10
Error code history, pos. 10
4
R
x
384
TempRotor
Temperature rotor
1
R
°C
0
999999
397
SetRotSpd
Set rotation speed (rpm)
1
R
rpm
0
999999
398
ActualSpd
Actual rotation speed (rpm)
1
R
rpm
0
999999
399
NominalSpd
Nominal rotation speed (rpm)
1
R
rpm
0
999999
#
Display
Designation
Functions
D
a
ta
ty
p
e
A
c
ce
ss
m
e
th
o
d
Unit
min max
default
#
Display
Designation
Functions
D
a
ta
ty
pe
A
cc
e
ss m
e
th
o
d
Unit
min max
default
700
RUTimeSVal
Set value run-up time
1 RW min
1
120
8
x
701
SpdSwPt1
Rotation speed switch point 1
1 RW %
50
97
80
x
707
SpdSVal
Set value in rot. speed setting mode
2 RW %
20
100
50
x
708
PwrSVal
Set value power consumption
7 RW %
10
100
100
1
x
710
Swoff BKP
Switching off threshold backing pump in in-
termittend mode
1 RW W
0
1000
0
x
711
SwOn BKP
Switching on threshold backing pump in in-
termittend mode
1 RW W
0
1000
0
x
717
StdbySVal
Set value rotation speed at standby
2 RW %
20
100
66.7
x
719
SpdSwPt2
Rotation speed switch point 2
1 RW %
5
97
20
x
720
VentSpd
Venting rot. speed at delayed venting
7 RW %
40
98
50
x
721
VentTime
Venting time at delayed venting
1 RW s
6
3600
3600
x
777
NomSpdConf
Nominal rotation speed confirmation
1 RW Hz
0
1500
x
797
RS485Adr
RS485 device address
1 RW
1
255
1
x
1. depending on the pump type
Option
Description
0 = Fan (continous operation)
Control via parameter Pumping station
1 = Venting valve, normally closed
Control via parameter Enable venting, when using a
venting valve which is normally closed.
2 = Heating
Control via parameters Heating and Rotation speed
switchpont attained
3 = Backing pump
Control via parameters Pumping station and operation
mode backing pump
4 = Fan (temperature controlled)
Control via parameter Pumping station and temperature
thresholds