The Pfeiffer Vacuum parameter set
olds for the backing pump are adjustable. Fluctuations in the power consumption
of idling turbopumps and type-dependent varying fore-vacuum pressures of the
backing pumps require the switching thresholds to be set separately for the inter-
mittend mode.
Pfeiffer Vacuum recommends the intermittend mode between 5 and 10 mbar. A
pressure gauge and a dosing valve are required to set the switching thresholds.
Switch on the vacuum system via the function "pumping station" and await the
run-up.
Generate a fore-vacuum pressure of 10 mbar by gas inlet via dosing valve.
Read and note the parameter
[P:316]
.
Adjust the switch on threshold backing pump via parameter
[P:711]
to the deter-
mined drive power for a fore-vacuum pressure of 10 mbar.
Reduce the fore-vacuum pressure to 5 mbar.
Read and note the parameter
[P:316]
.
Adjust the switch off threshold backing pump via parameter
[P:710]
to the deter-
mined drive power for a fore-vacuum pressure of 5 mbar.
Delayed switching on
Switching on the turbopump and the backing pump at the same time can result in
unwanted gas flows. Depending on process or application requirements, the back-
ing pump can be switched on with a delay. The switch-on delay depends on the ro-
tation speed of the turbopump and is fixed in the electronic drive unit at 6 Hz.
The switch-on delay signal can also be used for switching a fore-vacuum safety
valve.
Switching on the
pumping station
The function "pumping station" comprises turbopump operation with control of all
connected accessories (e.g. backing pump).
Switch on the supply voltage with switch S1 on the power supply.
Parameter
[P:023]
= 1
Parameter
[P:010]
= 1
Ongoing (and removed) error messages are reset. After a successfully completed
self-test, the electronic drive unit sets the turbopump motor and all connected ac-
cessories into operation depending on their configuration.
When the pumping station is activated, the turbopump can be switched off and on
via the function
[P:023]
.
Operation with acces-
sories
Depending on the configuration, various accessories can be connected to the tur-
bopump and controlled via parameter of the electronic drive unit.
Heating
Switch on or off the heating via parameter
[P:001]
.
The activation of a connected casing heating depends on rotation speed switch
point 1 (factory setting: 80 % x f
Nominal
).
Fan
Two options in the connection configuration enable continuous or temperature
controlled operation of a connected air cooling unit
. Threshold
values are type-specific and are anchored in the electronic drive unit.
Sealing gas valve
Switch on or off a sealing gas valve which is connected to a pre-configured out-
put via parameter
[P:050]
.