8
Operation
4.
Operation
Before operating the TIC for the first time the valid and unique serial interface address
has to be set by means of the Profibus address selection switch. Alterations to the
address during operations are not permissible.
Illuminating Displays
This LED provides information on the general unit status and the communication with
the TIC unit. Upon switching on there is a test phase of approximately one second for
the LED. Only thereafter will the
Table 4.0.1 on page 8
be valid.
Please note
☞
In the case of the TCP 350 (TIC 253) the time which elapses between switching on the
instrument until the Profibus is available may amount to approximately
15 seconds!
Status
Definition
Off
Unit is not supplied with adequate voltage
Green illuminating
Communication with the TIC unit running (status
data_exchange)
Green flashing
Bus traffic detected, no data_exchange
Green 1 x short, 1 x long flashing
The set address is not the one with which the unit is
currently working Abb. 2-1, Page 4
Red illuminating
Unit defective
Red flashing
No bus traffic detected
Red 1 x short, 1 x long flashing
Current parameter data are erroneous
Red 2 x short, 1 x long flashing
Current configuration data are erroneous
Tabelle 4.0.1: Illuminating Displays