Product Description

2.

Product Description

The TIC 253 respectively TIC 254 provides a connection for the bus system Profibus-DP 
(V0) for the drive units TCP 350 and TCP 3000 respectively TM 3000. If not stated other-
wise the designation "TIC" applies to the TIC 253 and also to the TIC 254.

Abb. 2-1 Description of the front panel

Connection options

The TIC offers connection options for the Profibus-DP by means of a 9-way Sub-D sok-
ket (TIC 253) respectively by means of two B-coded M12 connectors (TIC 254).

2.1

Delivery

Included with the delivery of the TIC 253/254 are the following items:

Unit TIC 253 respective TIC 254

CD with GSD file and graphics for inclusion in the Profibus configuration tool.

2.2

Proper Use

The TIC serves the purpose of connecting the aforementioned drive units to the Profi-
bus-DP (V0). The type of protection stated for the TIC 254 can only be attained in the 
case of correctly fitted rubber stoppers to the address selector switches and the service 
connection as well as with connected Profibus connectors, respectively connector with 
terminating resistor.

1

1

2

5

3

4

3

4

1

2

TIC 253

TIC 254

Profibus-DP connection

Status display

Profibus-DP Address 
(hexadecimal, most significant bit „msd“)

Profibus-DP Address 
(hexadecimal, least significant bit „lsd“)

Service connection

Содержание TIC 253

Страница 1: ...Betriebsanleitung Operating Instructions PT0138BE A 0507 TIC 253 TIC 254 Profibus DP gateway for TCP 350 and TCP TM 3000 ...

Страница 2: ...n 6 3 2 Profibus Connection 6 4 Operation 8 5 Modules 9 5 1 Module 1 standard 10 5 2 Module 2 power 11 5 3 Module 3 ophrs_pump 11 5 4 Module 4 ophrs_drv 11 5 5 Module 5 tms 12 5 6 Module 6 venting 12 5 7 Module 7 fixed_value_out 12 5 8 Module 8 fixed_value_in 13 5 9 Module 13 fail_safe 13 5 10 Module 14 acyclic access 14 6 Extended Diagnosis Data 15 7 What To Do In Case Of Breakdowns 15 8 Maintena...

Страница 3: ...ssembling accurate Do not carry out any unauthorised conversions or modifications on the unit Do not open the housing cover when the unit is connected to the mains nor during pumping operation Take account of the prescribed voltage when connecting the cable to the various plugs When returning the unit to us please note the shipping instructions 1 1 Pictogram Definitions 1 2 For Your Orientation Op...

Страница 4: ... of the TIC 253 254 are the following items Unit TIC 253 respective TIC 254 CD with GSD file and graphics for inclusion in the Profibus configuration tool 2 2 Proper Use The TIC serves the purpose of connecting the aforementioned drive units to the Profi bus DP V0 The type of protection stated for the TIC 254 can only be attained in the case of correctly fitted rubber stoppers to the address selec...

Страница 5: ...d above and in particular Connection to pumps and units which is not permitted in their operating instruc tions Connection to units which contain touchable and voltage carrying parts Improper use will cause any rights regarding liability and guarantees to be forfeited ...

Страница 6: ...ith your Profibus configuration tool and in the see Chap 5 page 9 3 2 Profibus Connection The Profibus is connected to this box The pin arrangement is in accordance with the Profibus standard Thereby the functions are supported in accordance with the pin assi gnment in the table below It is recommended to tighten the securing screws of the plug to prevent unintentional disconnection of the plug TI...

Страница 7: ...0 10 16 20 32 30 48 40 64 50 80 60 96 70 112 01 1 11 17 21 33 31 49 41 65 51 81 61 97 71 113 02 2 12 18 22 34 32 50 42 66 52 82 62 98 72 114 03 3 13 19 23 35 33 51 43 67 53 83 63 99 73 115 04 4 14 20 24 36 34 52 44 68 54 84 64 100 74 116 05 5 15 21 25 37 35 53 45 69 55 85 65 101 75 117 06 6 16 22 26 38 36 54 46 70 56 86 66 102 76 118 07 7 17 23 27 39 37 55 47 71 57 87 67 103 77 119 08 8 18 24 28 4...

Страница 8: ...se of the TCP 350 TIC 253 the time which elapses between switching on the instrument until the Profibus is available may amount to approximately 15 seconds Status Definition Off Unit is not supplied with adequate voltage Green illuminating Communication with the TIC unit running status data_exchange Green flashing Bus traffic detected no data_exchange Green 1 x short 1 x long flashing The set addr...

Страница 9: ...the input data of the modules a maximum of 42 parameters may be present Pxxx In the configuration no other module may be present after the fail_safe mo dules In order to update the input data as rapidly as possible please use only the mo dules you really require and do not use multiple parameter numbers Access to the functions in the modules can be influenced by external circuitry of the drive uni...

Страница 10: ...1 i Rotation speed switch point as a of the final rotation speed P701 Input data a Unit remote controlled P300 0 no 1 yes b Switchpoint attained P302 0 no 1 yes c Set rotation speed attained P306 0 no 1 yes d Pump accelerates P307 0 no 1 yes e TMS controller responded P333 0 no 1 yes f Actual rotation speed as a of the final rotation speed g Error Warning number low byte 0 no error 1 999 error 1 9...

Страница 11: ...s_drv This module may only be used a maximum of once Input data Configuration data 41h 83h 04h Byte 0 1 2 3 Current in 10 mA P310 Voltage in 10 mV P313 low byte high byte low byte high byte Byte 0 1 2 3 Operating hours of the pump P311 low word low byte low word high byte high word low byte high word high byte Byte 0 1 2 3 Operating hours of the electronic drive unit P314 low word low byte low wor...

Страница 12: ... described the parameter number of which is stated in the parameter data low byte first The output data must then contain a valid setpoint For possible parameters see Chap 11 page 17 Output data Configuration data 81h C0h 07h Parametering data Byte 0 Set value TMS heating in C P704 Byte 0 Actual value TMS heating in C P331 Byte 0 1 2 Venting frequency as a of the final rotation speed P720 Venting ...

Страница 13: ...mple switch pump off in case the Profibus connection fails In the parameter data the parameter number and the parameter value are stated which shall be written in this case As to the possible parameters see Chap 11 page 17 This module can be used up to four times in the configuration Configuration data 00h Parametering data Byte 0 1 Parameter value low byte high byte Byte 0 1 Parameter number low ...

Страница 14: ...the output data change Output data a Service e g read or write see Chap 11 page 17 b Path see Chap 11 page 17 d0 d3 Data don t care on read access format see Chap 11 page 17 Input data a Contains field a of output data with Bit 6 set e g 1 65 on correct execution on error value 71 b Path repeats path in output data d0 d7 Data don t care on write access format see Chap 11 page 17 Configuration data...

Страница 15: ...le 6 This module does not support the data type of the addressed parameter 7 The standard module is not located in the first connection slot 8 An error has been acknowledged although no error has occurred 9 The maximum number of permissible fail_safe modules has been exceeded 10 Unknown module 13 The maximum permissible number of the requested parameter numbers has been exceeded 14 Modules which m...

Страница 16: ... can provide full details 9 Technical Data 10 Accessories CAUTION The unit is not prepared for customer repair Units returned to us for repair or maintenance are covered by our general conditions of sale and supply Please refer to the back cover of this manual for contact addresses and telephone hotline numbers Description Unit TIC 253 TIC 254 Profibus version DP V0 Baud rates baud 9 6 k 19 2 k 45...

Страница 17: ...2 1 1 0 0 300 Unit remote controlled 1 1 9 2 18 byte 301 Oil deficiency TMP 1 1 3 1 17 word t 302 Rotation speed switch point attained 1 1 5 1 18 byte 303 Actual error code 1 1 3 1 18 word u 304 Over temperature drive unit 1 1 3 1 17 word t 305 Over temperature TMP BKP 1 1 3 1 17 word t 306 Set rotation speed attained 1 1 4 1 25 byte v 307 Pump accelerates 1 1 4 1 25 byte v 308 Set rotation speed ...

Страница 18: ...e 5 1 21 byte 701 Switchpoint in of final rotation speed 1 5 1 17 byte 704 TMS heating set value in C 5 6 2 20 word 707 Rotation speed set value in rot speed set ting mode 1n 4 1 24 real 710 Poff for VVP interval operation in W 6 1 21 word 711 Pon for VVP interval operation in W 6 1 22 word 717 Standby frequency 4 1 29 real 720 Venting frequency 6 6 4 22 dword 721 Venting time 6 6 4 20 word 777 Se...

Страница 19: ... in accordance with IEEE 754 4 bytes low byte first Bit 5 1 Insufficient oil 0 Sufficient oil Bit 3 1 Overtemperature 0 No overtemperature Field of the data type stated first byte contains the number of subsequent elements for example p349 field of byte 7 84 67 80 32 51 53 48 7 elements TCP 350 p360 field of word 1 6 0 1 element word 2 Byte value 6 0 stop 1 accelerating 2 at nominal speed 3 decele...