![Pfeiffer Vacuum TIC 253 Скачать руководство пользователя страница 6](http://html.mh-extra.com/html/pfeiffer-vacuum/tic-253/tic-253_operating-instructions-manual_1534949006.webp)
6
Installation
3.
Installation
3.1
Preparations For Installation
➡
Connect the Profibus cable to the TIC and secure it against inadvertent loosening by
screwing it down. When doing so, note the applicable regulations and recommenda-
tions for installing a Profibus system.
➡
Set the address selector switch to the Profibus address intended for this instru-
ment.
➡
TIC 254: fit the rubber stoppers to the address selector switches so as to attain the
type of protection stated. The stoppers must be inserted straight and as deeply as
possible into the holes.
➡
To configure the TIC use the enclosed GSD file. Information on this can be found
with your Profibus configuration tool and in the see Chap. 5., page 9.
3.2
Profibus Connection
The Profibus is connected to this box. The pin arrangement is in accordance with the
Profibus standard. Thereby the functions are supported in accordance with the pin assi-
gnment in the table below. It is recommended to tighten the securing screws of the plug
to prevent unintentional disconnection of the plug.
TIC 253
Pin
Function
1
Shield
2
-
3
RxD/TxD-P
4
CNTR-P
5
DGND
6
VP
7
-
8
RxD/TxD-N
9
-
TIC 254
Pin
Function
plug
socket
1
VP
2
RxD/TxD-N
3
DGND
4
RxD/TxD-P
5
Shield
Socket and plug are wired in
parallel so that they provide the
same functions.
For the Profibus connection of the TIC 254 you may use
the kits listed as accessories, see Chap. 10., page 16
Tabelle 3.2.1: Pin assignment Profibus connector
9
8
7
6
5
4
3
2
1
5
2 1
3 4
5
1 2
4 3