EN – 18
Reassembling
the seal-holder
(
see page N – 2
)
Reassembling the fan
and the motor side
components
(
see page N – 2
)
(
See page 7
)
.
Fill with oil
Fit the shaft key (
22
)
(N – 10)
on HP rotor.
Fit the coupling fan (
33
) and secure it with the screw (
28
) and washer (
26
).
Fit the drive key on the motor shaft. Install the motor coupling (
3
)
(N – 18)
down to
the stop on the motor shaft and secure it with the screw (
2
)
(N – 18)
.
Install the plastic coupling (
4
)
(N – 18)
on the motor coupling (
3
)
(N – 18)
.
Fit the motor on the frame and secure with the 4 mounting bolts (
5
)
(N – 18)
.
After reassembling, fill the pump with oil before start-up the pump (
see Operating
instructions
).
Reassembling of the
pumping pipe
(model SDI)
Position one the end of the pipe (
38
) into the housing hole designed for this purpose
and the other end on the stator (
34
).
Do not forget to install o-rings (
39
) and (
40
) after having checked their status (no cuts).
Reassembling the pump (cont’d)