Error
code
Problem
Possible causes
Remedy
Err098
Internal communication er-
ror
–
● Contact Pfeiffer Vacuum Service
Err114
Final stage temperature
evaluation faulty
–
● Contact Pfeiffer Vacuum Service
Err118
Excess temperature, final
stage
● Insufficient
cooling
● Check to make sure the fan is working.
● Check the fan connecting cable for a correct fit and
damage.
● Improve the cooling
● Check the operating conditions
Err175
PFC excess current
● Contact Pfeiffer Vacuum Service
Err176
Flash Program Error
● Contact Pfeiffer Vacuum Service
Err825
Internal communication er-
ror
● Contact Pfeiffer Vacuum Service
Err913
Direction of rotation
Incorrect direction
of rotation
● Contact Pfeiffer Vacuum Service
Tbl. 22:
Error messages of the electronic drive unit
Error
code
Problem
Possible causes
Remedy
Wrn007
Undervoltage or
power failure
● Mains failure
● Check the mains input voltage
Wrn037
Auto standby inac-
tive
● Control set to remote
● Deactivate the remote priority
Wrn038
Pressure regula-
tion error
● Target pressure can-
not be reached
● Increase or decrease the gas load
Wrn040
Note
Maintenance level 1 rec-
ommended
● Regularly check the performance data of the vacuum
pump
● If necessary, perform maintenance
● To reset the message, press
and
for 5 s at
the same time
Wrn042
Maintenance re-
quired
Have maintenance per-
formed
● Contact Pfeiffer Vacuum Service
Wrn044
Electronics high
temperature
● Insufficient cooling
● Check to make sure the fan is working.
● Check the fan connecting cable for a correct fit and
damage.
● Improve the cooling
● Check the operating conditions
Wrn045
Motor high temper-
ature
● Insufficient cooling
● Check to make sure the fan is working.
● Check the fan connecting cable for a correct fit and
damage.
● Improve the cooling
● Check the operating conditions
Wrn118
High power elec-
tronics temperature
● Insufficient cooling
● Check to make sure the fan is working.
● Check the fan connecting cable for a correct fit and
damage.
● Improve the cooling
● Check the operating conditions
Tbl. 23:
Warning messages of the electronic drive unit
Malfunctions
68/84
Содержание HISCROLL 12
Страница 52: ...Alternative Switch off via remote interface Remove the bridge between pins 7 ad 5 Operation 52 84...
Страница 83: ...83 84...