4. Secure the fan housing with both hexagon socket screws and washers.
– Tightening torque:
3.5 Nm
5.
Optional
: Fasten an existing mating plug on the “remote” connection of the electronic drive unit.
– Tightening torque:
0.4 Nm
9.6 Final inspection
Prerequisite
● Maintenance work carried out on opening the housing
Carrying out a final inspection
► Perform a functional test.
► Check to make sure the fan is working.
Recommended final check
► Carry out a leak test.
Maintenance
63/84