M4
Fig. 18:
Example: Connecting the grounding cable
Procedure
1. Use the turbopump ground terminal (M4 female thread).
2. Route the connection in accordance with locally applicable provisions.
5.6.2 Establishing the electric connection
DANGER
Danger to life from electric shock
Power supply packs that are not specified or are not approved will lead to severe injury to death.
► Make sure that the power supply pack meets the requirements for double isolation between
mains input voltage and output voltage, in accordance with IEC 61010-1 IEC 60950-1 and
IEC 62368-1.
► Make sure that the power supply pack meets the requirements in accordance with IEC 61010-1
IEC 60950-1 and IEC 62368-1.
► Where possible, use original power supply packs or only power supply packs that correspond
with the applicable safety regulations.
WARNING
Danger of cut injuries from unexpected start up.
The use of mating plugs of the electronic drive unit (accessories) enables the automatic run-up of the
vacuum pump as soon the power is turned on. Attaching mating plugs before or during the installa-
tion leads to the movement of parts hence the risk of cut injuries by sharp-edged in the exposed high
vacuum flange.
► Only connect mating plugs after the mechanical installation.
► Only switch on the vacuum pump immediately prior to operation.
Installation
33/62
Содержание HiPace 300 P
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original HIPACE 300 P Turbopump ...
Страница 54: ...13 Accessories View the range of accessories for hybrid bearing turbopumps on our website Accessories 54 62 ...
Страница 61: ...61 62 ...
Страница 62: ...Notizen Notes ...
Страница 63: ......
Страница 64: ... PT0633 ed C Date 2108 P N PT0633BEN ...