Type of gas or vapor pumped
Recommended
equipment
Actions/Monitoring to be
done
Minimum ACP configuration ac-
cording to versions
SD
G
CV
Inflammable
or explosive
gas
Work outside
the flammability
range of the
product (ideal =
25 to 50% of
the LEL
1
and/or
below the
MOC)
2
Connect the
gas ballast to a
neutral gas
source.
● Dilute the pumped gas
upstream or in the pump
to lower its concentration
below the lower flamma-
bility limit via purges
and/or the neutral gas
ballast.
● Dilute the gas dis-
charged by the pump to
lower its concentration to
25% of the LEL via purge
and/or the inert gas bal-
last.
● Avoid any build-up of gas
in the pumping line.
● Avoid air or humidity
backstreaming by main-
taining a gas speed
> 0.1 m/s in the exhaust
line.
● Check the sealing of the
installation.
-
Purge
open
Purge
open +
gas bal-
last open
1
Reference to USA NFPA 69-2002, chap. 6.3.1 Combustible concentration threshold; LEL = lower explosive limit
2
MOC Minimum Oxygen Concentration
Feature for CP - LG - SH pump versions
CP - LG - SH pump versions are designed for applications with specific qualities.
Contact our service center.
6.3 Different control modes
This chapter describes the connections and protocols associated with each control mode. 3 control
modes are available:
●
LOCAL
The pump is controlled by a switch:
I/O
. The pump runs independently of the equipment into
which it has been integrated.
●
REMOTE
The pump is remote controlled by opening or closing different dry contacts. Its rotation speed is
configured on the terminals of the remote control connector (see chapter “Remote mode opera-
tion”.
●
SERIAL LINK
The pump is controlled by commands sent via serial link
RS-485
(see chapter “Use in RS-485 link
mode”).
6.3.1 Local mode operation
In local mode,
the pump can run only if the cover plug
is fitted on the remote control connector. This
cover plug is delivered with the pump.
1
5
6
7
2
3
4
9
10
15
14
13
12
11
8
Fig. 6:
Cover plug with strap for operation in local mode
Operation
31/56
Содержание ACP 28 CP
Страница 1: ...OPERATING INSTRUCTIONS EN Translation of the Original ACP 28 40 Multi stage Roots pump air cooled...
Страница 54: ......
Страница 56: ...112369 Ed 09 Date 2018 06 P N 112369OEN...