34 . Spotlight 400 Series User’s Guide
Viewing in Reflectance/Diffuse Reflectance
When the imager is in viewing mode in reflectance or diffuse reflectance (Figure 11):
•
The reflectance illuminator assembly directs light from the upper LED down through one
side of the cassegrain onto the sample.
•
The upper cassegrain collects the reflected light from the sample and sends it upward
through the aperture.
The lower ‘transmission’ illuminator is also active in reflectance and diffuse reflectance
modes, giving simultaneous illumination from above and below.
Figure 11 Path of the visible beam for viewing a sample in reflectance/diffuse
reflectance
Mirror pair
Visible focus lens
assembly
Variable aperture
Reflectance
illuminator assembly
Upper cassegrain
Sample position
Lower cassegrain
Lower dichroic
mirror
Detector
Camera