108 . Spectrum Two N User's Guide
3.
Click
.
The Source Replacement Wizard is displayed.
4.
Follow the instructions on-screen.
5.
When you have completed the procedure, we recommend that you run the Component
Checks to confirm that your instrument is working correctly.
6.
Allow the source temperature to stabilize for approximately 2 hours before using the
instrument.
NOTE:
If you are using the AVI correction, you will need to run a new AVI calibration after
replacing the source.