no local analyte matching the Specimen Gate analyte, the plate will be disabled. Please contact your local PerkinElmer

representative or Specimen Gate support to solve the issue.
The Panthera-Puncher 9 can be used in two modes that define how it will get the sample queries. In Specimen Gate

mode (direct link to Specimen Gate Laboratory), the Panthera-Puncher 9 gets the sample queries from the Specimen

Gate Laboratory software to which it is connected. In stand-alone mode, it gets the sample queries from the analyte

settings and the loaded plates.

Note: 

If a sample does not fit the Specimen Gate barcode rules, you need to remove the specimen from the sample

plane and reintroduce it to refresh queries. If the barcode cannot be read from the sample, you can type in the barcode

before introducing the sample to the instrument.
When the plate is getting full, the instrument will display a pop-up warning if all replicates of an analyte punch query

cannot be punched to the same plate. This serves as a warning that unless the plate in question is substituted with a

new one, some queries will not be punched.
Changing a card before all its required disks have been punched will prompt the instrument to issue a warning.

The warning will not count queries that have been disabled (either manually or automatically by the Specimen Gate

Laboratory software).

Sample identification

If your laboratory uses barcode identification of samples, the software detects the barcode from the camera image and

assigns the correct ID to the sample. When using barcodes, make sure you can see the whole barcode on the screen

when inserting the card in order to allow the instrument to read the ID properly. Once the instrument has read the

barcode, you can adjust the sample card position as needed. Should a barcode be unreadable or missing or in need of

editing, the software lets you enter the ID manually. You will not be able to punch a sample unless it has been assigned

an ID. If you have an active sample barcode rule, the software will assign a query to the sample only if the barcode

follows the rule. For more details refer to 

Barcode rule for samples

 on page 24

If your laboratory is not using barcodes, the software includes an automatic identification system using running

numbers. This can be turned on or off from the 

Settings

 view (for more details see 

Automatic Sample ID generation

 on

page 24).
To provide the sample ID manually, press the 

Sample ID

 field and enter the details using the on-screen keyboard. The

sample ID must be at least two characters long.
Please bear in mind that the use of barcodes is highly recommended, as it ensures positive identification of the sample.

Both the automatic sample ID generation and manual sample ID entry allow for operator error and could lead to

patients being misidentified.
Calibrators and controls are also identified by their barcode. This means that all calibrators and controls must have

a barcode and a corresponding barcode rule defined in the analyte settings. If you have 

Automatic Sample ID

generation

 turned on, the instrument will not read barcodes, so you must switch it off in order to punch calibrators

and controls with barcodes.
You can also enter notes for a specific sample from the 

Punch

 view by pressing the 

Note

 field. When you enter a note

from the punch view, it is added to all wells punched subsequently from the same card until the card is removed.

Detection of punch areas

Adjust the sample card position so that the dried blood spot you want to punch fits as much as possible in the target

area marked with a circle. The software detects the areas where it can punch and displays each suggested punch as a

plain green disk with its number in it. Once the punch command is given, all the disks shown on the camera image will

be punched.

Note: 

The Panthera-Puncher 9 detects blood areas but does not detect spots as such, so it may suggest punching

required disks from two different spots if these are in the punching area (circle) defined by the camera.

Note: 

The minimum space to be left between two disks and between a disk and the blood spot border can be

configured through the service software.
When using 

Intellipunch

, the software will optimize the available surface area of the blood spot and will only show

on the screen the disks it can punch. If you are using a static punch pattern, any disk that cannot be punched

on a particular spot appears as a green circle (provided the 

Blood detect

 function is on in the 

Manual punch

configuration

 view. For more details on this, please refer to 

Punching patterns

 on page 14).

Panthera-Puncher

 9 User Manual

Operations in detail

13

Содержание Panthera-Puncher 9

Страница 1: ...2081 0010 USER MANUAL Panthera Puncher 9 Software version 1 8 Dried Blood Spot Punching Device...

Страница 2: ......

Страница 3: ...and setting up the plates 9 Unloading the plates from the instrument 10 Plate edit view 10 Operations in detail 12 Punching and sample queries 12 Sample identification 13 Detection of punch areas 13...

Страница 4: ...25 Loading templates 25 Analytes 25 Punching 26 Punches per well 26 Card barcode rules 27 Plate 27 Exporting 27 Plate maps 28 Barcode rule notation 28 Storing of the sample card image 30 Log files 31...

Страница 5: ...t from the Keyboards and Languages tab of the Windows Region and Language dialog The default user and administrator user names and corresponding passwords are provided in a separate document If the ne...

Страница 6: ...of moving parts in the instrument This way up Fragile handle with care Keep dry Certification mark from testing laboratory In vitro diagnostic medical device Catalogue number Battery recycle WEEE foll...

Страница 7: ...heck the number of disks in the well in error or repunch situations Incorrect number of disks may lead to a false result Warning Installing the instrument in a location where sunlight reaches the samp...

Страница 8: ...eads Caution When cleaning the punch heads ensure that parts from different punch heads do not become interchanged This may cause the punch head to become unusable the disk to not have space to fall o...

Страница 9: ...need to be added to certain user groups in order to have access rights to the Panthera software User groups in question are Puncher Users or Puncher Administrators depending on what functions the use...

Страница 10: ...nched from each blood spot Intellipunch detects the blood areas on the sample card based on the card surface color and automatically optimizes the punch locations for the current blood spot This provi...

Страница 11: ...t able to recover the error message will say Unrecoverable error When the instrument informs the user of e g a movement error it is a hint that an error has occurred but that the instrument was able t...

Страница 12: ...all controls 6 green dots to the plate prior to patient samples By default it is set to punch only the start controls so that the plate can be packed in case it is a partial plate It is not possible...

Страница 13: ...cause the plates to become unusable Loading and setting up the plates Press the Eject button in the Plates view and open the hatch Once the hatch is open the plate track comes out The plate track cont...

Страница 14: ...head configuration When connected to Specimen Gate Laboratory all samples with the unavailable disk size will still be displayed on the punch screen albeit they will be disabled and you will not be a...

Страница 15: ...now want to export data the default folder path is that of the file previously imported If needed remember to change the folder path to point to the correct location If a plate needs to be reloaded to...

Страница 16: ...er the first punch command realign the card and the instrument will locate the next positions to punch the missing disks Tip You can easily toggle from the Plates view to the Punch view by bringing th...

Страница 17: ...utomatic identification system using running numbers This can be turned on or off from the Settings view for more details see Automatic Sample ID generation on page 24 To provide the sample ID manuall...

Страница 18: ...osition In Specimen Gate mode the right punch head is always brought to the front while waiting for the punch command as the instrument does not know the correct prioritized size before the queries ar...

Страница 19: ...ears red and each plate with a red well is outlined in red as well Select each plate outlined in red to see the detailed view Once you have visually checked the well you can change its status by selec...

Страница 20: ...ckground of the Punch head maintenance area also appears in red Even if the software reminds you to clean the punch heads you are be able to continue punching However it is strongly advised that you p...

Страница 21: ...emoving and cleaning the punch heads In order to clean a punch head you need to remove it from the puncher dismantle it and then clean the various parts The inner chute is the part that gathers the mo...

Страница 22: ...tor 90 degrees counterclockwise and slide it out The disk detector the chute and the spring come out of the punch head Remove the chute and the spring by pulling them Panthera Puncher 9 User Manual Ma...

Страница 23: ...Separate the spring from the chute In principle you should have the different components as shown in the picture below Panthera Puncher 9 User Manual Maintenance and shutdown 19...

Страница 24: ...ater first should largely eliminate this problem You may use swabs foam or nylon applicator sticks are best as cotton tipped applicator sticks can leave behind cotton filaments or cleaning clothes tha...

Страница 25: ...at a 90 degrees angle then insert the inner chute part into the guide and make sure that the fastening pins of the frame arm are placed in the groove of the plastic part of the detector You can then...

Страница 26: ...position and replace the punch head into the instrument Note On occasion when inserting the punch head back to the instrument after the daily punch head maintenance the Panthera software presents a Sk...

Страница 27: ...It is critical that you protect the detector from any liquid Panthera Puncher 9 User Manual Maintenance and shutdown 23...

Страница 28: ...0531 yy M d 11 5 31 MM dd BB 05 31BB AB yy ooo where ooo represents an ordinal date i e the number corresponding to the current day in the current year the 31st of May is the 151st day of the year 201...

Страница 29: ...ap autoincrement and start location parameters After pressing the button a new window opens and gives you the possibility to select a platemap to preview the platemap to set the start location for tha...

Страница 30: ...for a particular analyte press the corresponding button A page displays the analyte settings which are organized in six areas Punching Punches per well Card barcode rules Plate Plate maps Exporting Pu...

Страница 31: ...the rules for automatic identification of an analyte from a plate barcode See Card barcode rules on page 27 for basic instructions for writing rules For example if you want a rule for a plate barcode...

Страница 32: ...plate maps for the selected analyte Only three plate maps are previewed at a time You can browse through all plate maps by pressing the left and right arrows Every analyte must have at least one plat...

Страница 33: ...here x is the number of characters All static components in the barcode rules are written without The rules are case sensitive on the static components Table 1 Examples of barcode notation Rule Notati...

Страница 34: ...the sample cards punched on the corresponding day The name of each file is structured as follows xxxxxxxxx date time where xxxxxxxxx is of variable length and represents the sample card ID This is to...

Страница 35: ...are stored in the folder C ProgramData PerkinElmer SpotLight Logs The programdata folder is a hidden folder It can be accessed by typing the location in the address bar of Windows explorer Panthera Pu...

Страница 36: ...one more disk be punched from that same spot Is this normal This is due to the inaccuracy of pixel size calibration which ensures that the disks fit In some situations a disk can be a pixel smaller th...

Страница 37: ...about this product or you need assistance please contact PerkinElmer Customer Support Rest of the world GS customersupport perkinelmer com North America gs technical support perkinelmer com Panthera...

Страница 38: ...number listed below Give the full identification number of the manual you require This number is at the top right hand corner of the first page of the PDF The format is 2081 9xxx xx Each language ver...

Страница 39: ...ation 16 Maintenance and shutdown 16 O Operation Loading and setting up plates 9 Plate edit view 10 Punching and sample queries 12 Startup and shutdown 8 Operation continued Uloading plates from instr...

Страница 40: ...36...

Страница 41: ......

Страница 42: ...2081 9020 13 April 26 2022 Manufacturer Wallac Oy Mustionkatu 6 FI 20750 Turku Finland Tel 358 2 2678 111 Fax 358 2 2678 357 www perkinelmer com...

Отзывы: